Abstract

Oxidized zirconium, a material with a ceramic surface on a metal substrate, and highly cross-linked polyethylene are two materials developed to reduce wear. We measured in vivo femoral head penetration in patients with these advanced bearings. We hypothesized the linear wear rates would be lower than those published for cobalt-chrome and standard polyethylene. We retrospectively reviewed a select series of 56 THAs in a relatively young, active patient population utilizing oxidized zirconium femoral heads and highly cross-linked polyethylene acetabular liners. Femoral head penetration was determined using the Martell computerized edge-detection method. All patients were available for 2-year clinical and radiographic followup. True linear wear was 4 μm/year (95% confidence intervals, ± 59 μm/year). The early wear rates in this cohort of relatively young, active patients were low and we believe justify the continued study of these alternative bearing surfaces. Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.

Original languageEnglish (US)
Pages (from-to)141-145
Number of pages5
JournalClinical Orthopaedics and Related Research
Volume467
Issue number1
DOIs
StatePublished - Jan 2009

Fingerprint

Polyethylene
Thigh
Tacrine
Ceramics
Cobalt
Metals
Guidelines
Confidence Intervals
Population
Therapeutics

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Cite this

Wear analysis in tha utilizing oxidized zirconium and crosslinked polyethylene. / Garvin, Kevin Lloyd; Hartman, Curtis W; Mangla, Jimmi; Murdoch, Nathan; Martell, John M.

In: Clinical Orthopaedics and Related Research, Vol. 467, No. 1, 01.2009, p. 141-145.

Research output: Contribution to journalArticle

@article{27f5b85a658846348909160d71010967,
title = "Wear analysis in tha utilizing oxidized zirconium and crosslinked polyethylene",
abstract = "Oxidized zirconium, a material with a ceramic surface on a metal substrate, and highly cross-linked polyethylene are two materials developed to reduce wear. We measured in vivo femoral head penetration in patients with these advanced bearings. We hypothesized the linear wear rates would be lower than those published for cobalt-chrome and standard polyethylene. We retrospectively reviewed a select series of 56 THAs in a relatively young, active patient population utilizing oxidized zirconium femoral heads and highly cross-linked polyethylene acetabular liners. Femoral head penetration was determined using the Martell computerized edge-detection method. All patients were available for 2-year clinical and radiographic followup. True linear wear was 4 μm/year (95{\%} confidence intervals, ± 59 μm/year). The early wear rates in this cohort of relatively young, active patients were low and we believe justify the continued study of these alternative bearing surfaces. Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.",
author = "Garvin, {Kevin Lloyd} and Hartman, {Curtis W} and Jimmi Mangla and Nathan Murdoch and Martell, {John M.}",
year = "2009",
month = "1",
doi = "10.1007/s11999-008-0544-5",
language = "English (US)",
volume = "467",
pages = "141--145",
journal = "Clinical Orthopaedics and Related Research",
issn = "0009-921X",
publisher = "Springer New York",
number = "1",

}

TY - JOUR

T1 - Wear analysis in tha utilizing oxidized zirconium and crosslinked polyethylene

AU - Garvin, Kevin Lloyd

AU - Hartman, Curtis W

AU - Mangla, Jimmi

AU - Murdoch, Nathan

AU - Martell, John M.

PY - 2009/1

Y1 - 2009/1

N2 - Oxidized zirconium, a material with a ceramic surface on a metal substrate, and highly cross-linked polyethylene are two materials developed to reduce wear. We measured in vivo femoral head penetration in patients with these advanced bearings. We hypothesized the linear wear rates would be lower than those published for cobalt-chrome and standard polyethylene. We retrospectively reviewed a select series of 56 THAs in a relatively young, active patient population utilizing oxidized zirconium femoral heads and highly cross-linked polyethylene acetabular liners. Femoral head penetration was determined using the Martell computerized edge-detection method. All patients were available for 2-year clinical and radiographic followup. True linear wear was 4 μm/year (95% confidence intervals, ± 59 μm/year). The early wear rates in this cohort of relatively young, active patients were low and we believe justify the continued study of these alternative bearing surfaces. Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.

AB - Oxidized zirconium, a material with a ceramic surface on a metal substrate, and highly cross-linked polyethylene are two materials developed to reduce wear. We measured in vivo femoral head penetration in patients with these advanced bearings. We hypothesized the linear wear rates would be lower than those published for cobalt-chrome and standard polyethylene. We retrospectively reviewed a select series of 56 THAs in a relatively young, active patient population utilizing oxidized zirconium femoral heads and highly cross-linked polyethylene acetabular liners. Femoral head penetration was determined using the Martell computerized edge-detection method. All patients were available for 2-year clinical and radiographic followup. True linear wear was 4 μm/year (95% confidence intervals, ± 59 μm/year). The early wear rates in this cohort of relatively young, active patients were low and we believe justify the continued study of these alternative bearing surfaces. Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.

UR - http://www.scopus.com/inward/record.url?scp=58249085893&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=58249085893&partnerID=8YFLogxK

U2 - 10.1007/s11999-008-0544-5

DO - 10.1007/s11999-008-0544-5

M3 - Article

C2 - 18946711

AN - SCOPUS:58249085893

VL - 467

SP - 141

EP - 145

JO - Clinical Orthopaedics and Related Research

JF - Clinical Orthopaedics and Related Research

SN - 0009-921X

IS - 1

ER -