TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis

S. Daibes Figueroa, C. Winkelmann, W. Miller, T. Rold, G. Sieckman, C. Smith, L. Ma, Jered C Garrison, W. Volkert, T. Hoffman

Research output: Contribution to journalArticle

Abstract

Purpose: Because Micro‐CT utilizes ionizing radiation for image formation, radiation exposure during imaging is a concern. The objective of this study is to quantify the radiation exposure delivered during a Micro‐CT scan and to assess potential therapeutic effects associated with this radiation dose in a murine cancer model. Materials and Methods: Radiation exposure was measured using calibrated thermoluminescent dosimeters (TLD‐100) irradiated during a typical Micro‐CT scan protocol. TLD calibration curves were generated with a Cs‐137 irradiator. TLD's were implanted into a euthanized mouse and was imaged with Micro‐CT. TLD's were removed post‐scan and analyzed. Internal exposures were converted to dose in water. A C57BL/6 mouse lung tumor model derived by IV injection of 400,000 B16F10 murine melanoma cells was assessed for survivability and potential therapeutic effects due to absorbed radiation doses during Micro‐CT imaging. Results: A single Micro‐CT scan dose of 7.8±0.5 cGy was achieved when using a lucite anesthesia support module and a dose of 9.2±0.6 cGy with out the use of the anesthesia module. TLD data was validated using an ion chamber, providing measured radiation exposures of 8.1±0.4 cGy and 9.7±0.5 cGy with and with out the anesthesia module, respectively. Internal TLD analysis demonstrated an average mouse organ absorbed dose of 7.3±0.6 cGy. Conclusions: Survival analysis demonstrated a mean survival of nontreated control animals of 29±2 days, with animals receiving up to five sequential Micro‐CT studies surviving a mean of 30.5±1.5 days (total estimated dose of 39±2.5 cGy). The calculated cell survival fraction for a 9.2 cGy Micro‐CT scan was 99.25%. Therefore, negligible therapeutic effect from the radiation exposure delivered during Micro‐CT analysis was observed in the animal model investigated.

Original languageEnglish (US)
Pages (from-to)2211
Number of pages1
JournalMedical Physics
Volume33
Issue number6
DOIs
StatePublished - 2006
Externally publishedYes

Fingerprint

X-Ray Microtomography
Therapeutic Uses
Anesthesia
Radiation
Polymethyl Methacrylate
Survival Analysis
Ionizing Radiation
Inbred C57BL Mouse
Calibration
Melanoma
Neoplasms
Cell Survival
Animal Models
Radiation Exposure
Ions
Lung
Injections
Water

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Daibes Figueroa, S., Winkelmann, C., Miller, W., Rold, T., Sieckman, G., Smith, C., ... Hoffman, T. (2006). TU‐E‐330D‐01: TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis. Medical Physics, 33(6), 2211. https://doi.org/10.1118/1.2241612

TU‐E‐330D‐01 : TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis. / Daibes Figueroa, S.; Winkelmann, C.; Miller, W.; Rold, T.; Sieckman, G.; Smith, C.; Ma, L.; Garrison, Jered C; Volkert, W.; Hoffman, T.

In: Medical Physics, Vol. 33, No. 6, 2006, p. 2211.

Research output: Contribution to journalArticle

Daibes Figueroa, S, Winkelmann, C, Miller, W, Rold, T, Sieckman, G, Smith, C, Ma, L, Garrison, JC, Volkert, W & Hoffman, T 2006, 'TU‐E‐330D‐01: TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis', Medical Physics, vol. 33, no. 6, pp. 2211. https://doi.org/10.1118/1.2241612
Daibes Figueroa, S. ; Winkelmann, C. ; Miller, W. ; Rold, T. ; Sieckman, G. ; Smith, C. ; Ma, L. ; Garrison, Jered C ; Volkert, W. ; Hoffman, T. / TU‐E‐330D‐01 : TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis. In: Medical Physics. 2006 ; Vol. 33, No. 6. pp. 2211.
@article{aeb93cae87bb46948d971bcb82aaba28,
title = "TU‐E‐330D‐01: TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis",
abstract = "Purpose: Because Micro‐CT utilizes ionizing radiation for image formation, radiation exposure during imaging is a concern. The objective of this study is to quantify the radiation exposure delivered during a Micro‐CT scan and to assess potential therapeutic effects associated with this radiation dose in a murine cancer model. Materials and Methods: Radiation exposure was measured using calibrated thermoluminescent dosimeters (TLD‐100) irradiated during a typical Micro‐CT scan protocol. TLD calibration curves were generated with a Cs‐137 irradiator. TLD's were implanted into a euthanized mouse and was imaged with Micro‐CT. TLD's were removed post‐scan and analyzed. Internal exposures were converted to dose in water. A C57BL/6 mouse lung tumor model derived by IV injection of 400,000 B16F10 murine melanoma cells was assessed for survivability and potential therapeutic effects due to absorbed radiation doses during Micro‐CT imaging. Results: A single Micro‐CT scan dose of 7.8±0.5 cGy was achieved when using a lucite anesthesia support module and a dose of 9.2±0.6 cGy with out the use of the anesthesia module. TLD data was validated using an ion chamber, providing measured radiation exposures of 8.1±0.4 cGy and 9.7±0.5 cGy with and with out the anesthesia module, respectively. Internal TLD analysis demonstrated an average mouse organ absorbed dose of 7.3±0.6 cGy. Conclusions: Survival analysis demonstrated a mean survival of nontreated control animals of 29±2 days, with animals receiving up to five sequential Micro‐CT studies surviving a mean of 30.5±1.5 days (total estimated dose of 39±2.5 cGy). The calculated cell survival fraction for a 9.2 cGy Micro‐CT scan was 99.25{\%}. Therefore, negligible therapeutic effect from the radiation exposure delivered during Micro‐CT analysis was observed in the animal model investigated.",
author = "{Daibes Figueroa}, S. and C. Winkelmann and W. Miller and T. Rold and G. Sieckman and C. Smith and L. Ma and Garrison, {Jered C} and W. Volkert and T. Hoffman",
year = "2006",
doi = "10.1118/1.2241612",
language = "English (US)",
volume = "33",
pages = "2211",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - TU‐E‐330D‐01

T2 - TLD‐100 Measurement and Assessment of Internal Mouse Dosimetry During Micro‐CT Analysis

AU - Daibes Figueroa, S.

AU - Winkelmann, C.

AU - Miller, W.

AU - Rold, T.

AU - Sieckman, G.

AU - Smith, C.

AU - Ma, L.

AU - Garrison, Jered C

AU - Volkert, W.

AU - Hoffman, T.

PY - 2006

Y1 - 2006

N2 - Purpose: Because Micro‐CT utilizes ionizing radiation for image formation, radiation exposure during imaging is a concern. The objective of this study is to quantify the radiation exposure delivered during a Micro‐CT scan and to assess potential therapeutic effects associated with this radiation dose in a murine cancer model. Materials and Methods: Radiation exposure was measured using calibrated thermoluminescent dosimeters (TLD‐100) irradiated during a typical Micro‐CT scan protocol. TLD calibration curves were generated with a Cs‐137 irradiator. TLD's were implanted into a euthanized mouse and was imaged with Micro‐CT. TLD's were removed post‐scan and analyzed. Internal exposures were converted to dose in water. A C57BL/6 mouse lung tumor model derived by IV injection of 400,000 B16F10 murine melanoma cells was assessed for survivability and potential therapeutic effects due to absorbed radiation doses during Micro‐CT imaging. Results: A single Micro‐CT scan dose of 7.8±0.5 cGy was achieved when using a lucite anesthesia support module and a dose of 9.2±0.6 cGy with out the use of the anesthesia module. TLD data was validated using an ion chamber, providing measured radiation exposures of 8.1±0.4 cGy and 9.7±0.5 cGy with and with out the anesthesia module, respectively. Internal TLD analysis demonstrated an average mouse organ absorbed dose of 7.3±0.6 cGy. Conclusions: Survival analysis demonstrated a mean survival of nontreated control animals of 29±2 days, with animals receiving up to five sequential Micro‐CT studies surviving a mean of 30.5±1.5 days (total estimated dose of 39±2.5 cGy). The calculated cell survival fraction for a 9.2 cGy Micro‐CT scan was 99.25%. Therefore, negligible therapeutic effect from the radiation exposure delivered during Micro‐CT analysis was observed in the animal model investigated.

AB - Purpose: Because Micro‐CT utilizes ionizing radiation for image formation, radiation exposure during imaging is a concern. The objective of this study is to quantify the radiation exposure delivered during a Micro‐CT scan and to assess potential therapeutic effects associated with this radiation dose in a murine cancer model. Materials and Methods: Radiation exposure was measured using calibrated thermoluminescent dosimeters (TLD‐100) irradiated during a typical Micro‐CT scan protocol. TLD calibration curves were generated with a Cs‐137 irradiator. TLD's were implanted into a euthanized mouse and was imaged with Micro‐CT. TLD's were removed post‐scan and analyzed. Internal exposures were converted to dose in water. A C57BL/6 mouse lung tumor model derived by IV injection of 400,000 B16F10 murine melanoma cells was assessed for survivability and potential therapeutic effects due to absorbed radiation doses during Micro‐CT imaging. Results: A single Micro‐CT scan dose of 7.8±0.5 cGy was achieved when using a lucite anesthesia support module and a dose of 9.2±0.6 cGy with out the use of the anesthesia module. TLD data was validated using an ion chamber, providing measured radiation exposures of 8.1±0.4 cGy and 9.7±0.5 cGy with and with out the anesthesia module, respectively. Internal TLD analysis demonstrated an average mouse organ absorbed dose of 7.3±0.6 cGy. Conclusions: Survival analysis demonstrated a mean survival of nontreated control animals of 29±2 days, with animals receiving up to five sequential Micro‐CT studies surviving a mean of 30.5±1.5 days (total estimated dose of 39±2.5 cGy). The calculated cell survival fraction for a 9.2 cGy Micro‐CT scan was 99.25%. Therefore, negligible therapeutic effect from the radiation exposure delivered during Micro‐CT analysis was observed in the animal model investigated.

UR - http://www.scopus.com/inward/record.url?scp=85024792349&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024792349&partnerID=8YFLogxK

U2 - 10.1118/1.2241612

DO - 10.1118/1.2241612

M3 - Article

VL - 33

SP - 2211

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -