This study examined the theory and behavior of a chromatographic competitive binding immunoassay with sequential injection of sample and a labeled analyte analog. Based on nonlinear chromatographic theory, an equation was derived to describe the calibration curve for this assay in a system with adsorption-limited kinetics and homogeneous binding sites. This equation related assay response (B/Bo) to the column's binding capacity, the moles of analyte and labeled analog injected, and the flow rate/adsorption kinetics of the system. There was good agreement between this equation and experimental data for the binding of human serum albumin (HSA) to an immobilized anti-HSA antibody column. It was found that the amount of labeled analog injected, when applied in excess vs binding sites in the column, had little or no effect on the relative response or position of the calibration curve. The position of the curve was determined mainly by the number of binding sites on the column; however, this position could be shifted over several orders of magnitude by varying the flow rate used for analyte injection. By varying both the injection flow rate and labeled analog load, this method could be designed for use as either a screening test or quantitative assay.

Original languageEnglish (US)
Pages (from-to)1622-1630
Number of pages9
JournalAnalytical chemistry
Volume65
Issue number11
DOIs
StatePublished - Jun 1 1993

Fingerprint

Chromatography
Assays
Binding Sites
Flow rate
Serum Albumin
Calibration
Adsorption
Kinetics
Screening
Antibodies

ASJC Scopus subject areas

  • Analytical Chemistry

Cite this

Theory of a Sequential Addition Competitive Binding Immunoassay Based on High-Performance Immunoaffinity Chromatography. / Hage, David S.; Thomas, David H.; Beck, Melissa S.

In: Analytical chemistry, Vol. 65, No. 11, 01.06.1993, p. 1622-1630.

Research output: Contribution to journalArticle

@article{e253fc644cf64c0dac02ca7140ca85cd,
title = "Theory of a Sequential Addition Competitive Binding Immunoassay Based on High-Performance Immunoaffinity Chromatography",
abstract = "This study examined the theory and behavior of a chromatographic competitive binding immunoassay with sequential injection of sample and a labeled analyte analog. Based on nonlinear chromatographic theory, an equation was derived to describe the calibration curve for this assay in a system with adsorption-limited kinetics and homogeneous binding sites. This equation related assay response (B/Bo) to the column's binding capacity, the moles of analyte and labeled analog injected, and the flow rate/adsorption kinetics of the system. There was good agreement between this equation and experimental data for the binding of human serum albumin (HSA) to an immobilized anti-HSA antibody column. It was found that the amount of labeled analog injected, when applied in excess vs binding sites in the column, had little or no effect on the relative response or position of the calibration curve. The position of the curve was determined mainly by the number of binding sites on the column; however, this position could be shifted over several orders of magnitude by varying the flow rate used for analyte injection. By varying both the injection flow rate and labeled analog load, this method could be designed for use as either a screening test or quantitative assay.",
author = "Hage, {David S.} and Thomas, {David H.} and Beck, {Melissa S.}",
year = "1993",
month = "6",
day = "1",
doi = "10.1021/ac00059a023",
language = "English (US)",
volume = "65",
pages = "1622--1630",
journal = "Analytical Chemistry",
issn = "0003-2700",
publisher = "American Chemical Society",
number = "11",

}

TY - JOUR

T1 - Theory of a Sequential Addition Competitive Binding Immunoassay Based on High-Performance Immunoaffinity Chromatography

AU - Hage, David S.

AU - Thomas, David H.

AU - Beck, Melissa S.

PY - 1993/6/1

Y1 - 1993/6/1

N2 - This study examined the theory and behavior of a chromatographic competitive binding immunoassay with sequential injection of sample and a labeled analyte analog. Based on nonlinear chromatographic theory, an equation was derived to describe the calibration curve for this assay in a system with adsorption-limited kinetics and homogeneous binding sites. This equation related assay response (B/Bo) to the column's binding capacity, the moles of analyte and labeled analog injected, and the flow rate/adsorption kinetics of the system. There was good agreement between this equation and experimental data for the binding of human serum albumin (HSA) to an immobilized anti-HSA antibody column. It was found that the amount of labeled analog injected, when applied in excess vs binding sites in the column, had little or no effect on the relative response or position of the calibration curve. The position of the curve was determined mainly by the number of binding sites on the column; however, this position could be shifted over several orders of magnitude by varying the flow rate used for analyte injection. By varying both the injection flow rate and labeled analog load, this method could be designed for use as either a screening test or quantitative assay.

AB - This study examined the theory and behavior of a chromatographic competitive binding immunoassay with sequential injection of sample and a labeled analyte analog. Based on nonlinear chromatographic theory, an equation was derived to describe the calibration curve for this assay in a system with adsorption-limited kinetics and homogeneous binding sites. This equation related assay response (B/Bo) to the column's binding capacity, the moles of analyte and labeled analog injected, and the flow rate/adsorption kinetics of the system. There was good agreement between this equation and experimental data for the binding of human serum albumin (HSA) to an immobilized anti-HSA antibody column. It was found that the amount of labeled analog injected, when applied in excess vs binding sites in the column, had little or no effect on the relative response or position of the calibration curve. The position of the curve was determined mainly by the number of binding sites on the column; however, this position could be shifted over several orders of magnitude by varying the flow rate used for analyte injection. By varying both the injection flow rate and labeled analog load, this method could be designed for use as either a screening test or quantitative assay.

UR - http://www.scopus.com/inward/record.url?scp=0027617019&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027617019&partnerID=8YFLogxK

U2 - 10.1021/ac00059a023

DO - 10.1021/ac00059a023

M3 - Article

C2 - 8328676

AN - SCOPUS:0027617019

VL - 65

SP - 1622

EP - 1630

JO - Analytical Chemistry

JF - Analytical Chemistry

SN - 0003-2700

IS - 11

ER -