The Impact of Wire Stent Fabrication Technique on the Performance of Stent Placement

Shijia Zhao, Xiangyi Cheryl Liu, Linxia Gu

Research output: Contribution to journalArticle

19 Citations (Scopus)

Abstract

Braided wire stents demonstrate distinct characteristics compared to welded ones. In this study, both braided and welded wire stents with the same nominal dimensions were crimped inside a sheath and then deployed into a stenosed artery using finite element analysis. The braided wire stent was generated by overlapping wires to form crisscross shape. A welded wire stent was created by welding the intersection points of wires to avoid sliding between wires. The effect of fabrication technique on mechanical behavior of Nitinol wire stents was evaluated. The results showed that relative sliding between wires reduced the deformation of the braided stent, which led to less radial strength than the welded one; therefore, the deployed braided stent was more conformed to the anatomic shape of the lesion and much less efficient for restoring the patency of the stenotic artery. Post balloon-dilation was commonly used to improve its performance in terms of lumen gain and deployed shape of the stent. On the contrary, the welded wire stent exhibited a high capacity for pushing the occlusion outward. It reached an approximately uniform shape after deployment. The welded joints caused larger deformation and high strain on the stent struts, which indicate a potential earlier failure for the welded stent. In addition, higher contact pressure at the stent-lesion interface and higher arterial stresses were observed in the artery supported by the welded stent. The peak stress concentration may increase the occurrence of neointimal hyperplasia.

Original languageEnglish (US)
Article number011007
JournalJournal of Medical Devices, Transactions of the ASME
Volume6
Issue number1
DOIs
StatePublished - Mar 12 2012

Fingerprint

Stents
Wire
Fabrication
Arteries
Welding
Finite Element Analysis
Struts
Balloons
Hyperplasia
Stress concentration
Dilatation
Welds

Keywords

  • Braiding
  • Finite element method
  • Nitinol stent
  • Post balloon-dilation
  • Strain energy
  • Stress analysis
  • Welding

ASJC Scopus subject areas

  • Medicine (miscellaneous)
  • Biomedical Engineering

Cite this

The Impact of Wire Stent Fabrication Technique on the Performance of Stent Placement. / Zhao, Shijia; Liu, Xiangyi Cheryl; Gu, Linxia.

In: Journal of Medical Devices, Transactions of the ASME, Vol. 6, No. 1, 011007, 12.03.2012.

Research output: Contribution to journalArticle

@article{93c1844be81b4537864c4a789248f98a,
title = "The Impact of Wire Stent Fabrication Technique on the Performance of Stent Placement",
abstract = "Braided wire stents demonstrate distinct characteristics compared to welded ones. In this study, both braided and welded wire stents with the same nominal dimensions were crimped inside a sheath and then deployed into a stenosed artery using finite element analysis. The braided wire stent was generated by overlapping wires to form crisscross shape. A welded wire stent was created by welding the intersection points of wires to avoid sliding between wires. The effect of fabrication technique on mechanical behavior of Nitinol wire stents was evaluated. The results showed that relative sliding between wires reduced the deformation of the braided stent, which led to less radial strength than the welded one; therefore, the deployed braided stent was more conformed to the anatomic shape of the lesion and much less efficient for restoring the patency of the stenotic artery. Post balloon-dilation was commonly used to improve its performance in terms of lumen gain and deployed shape of the stent. On the contrary, the welded wire stent exhibited a high capacity for pushing the occlusion outward. It reached an approximately uniform shape after deployment. The welded joints caused larger deformation and high strain on the stent struts, which indicate a potential earlier failure for the welded stent. In addition, higher contact pressure at the stent-lesion interface and higher arterial stresses were observed in the artery supported by the welded stent. The peak stress concentration may increase the occurrence of neointimal hyperplasia.",
keywords = "Braiding, Finite element method, Nitinol stent, Post balloon-dilation, Strain energy, Stress analysis, Welding",
author = "Shijia Zhao and Liu, {Xiangyi Cheryl} and Linxia Gu",
year = "2012",
month = "3",
day = "12",
doi = "10.1115/1.4005788",
language = "English (US)",
volume = "6",
journal = "Journal of Medical Devices, Transactions of the ASME",
issn = "1932-6181",
publisher = "American Society of Mechanical Engineers(ASME)",
number = "1",

}

TY - JOUR

T1 - The Impact of Wire Stent Fabrication Technique on the Performance of Stent Placement

AU - Zhao, Shijia

AU - Liu, Xiangyi Cheryl

AU - Gu, Linxia

PY - 2012/3/12

Y1 - 2012/3/12

N2 - Braided wire stents demonstrate distinct characteristics compared to welded ones. In this study, both braided and welded wire stents with the same nominal dimensions were crimped inside a sheath and then deployed into a stenosed artery using finite element analysis. The braided wire stent was generated by overlapping wires to form crisscross shape. A welded wire stent was created by welding the intersection points of wires to avoid sliding between wires. The effect of fabrication technique on mechanical behavior of Nitinol wire stents was evaluated. The results showed that relative sliding between wires reduced the deformation of the braided stent, which led to less radial strength than the welded one; therefore, the deployed braided stent was more conformed to the anatomic shape of the lesion and much less efficient for restoring the patency of the stenotic artery. Post balloon-dilation was commonly used to improve its performance in terms of lumen gain and deployed shape of the stent. On the contrary, the welded wire stent exhibited a high capacity for pushing the occlusion outward. It reached an approximately uniform shape after deployment. The welded joints caused larger deformation and high strain on the stent struts, which indicate a potential earlier failure for the welded stent. In addition, higher contact pressure at the stent-lesion interface and higher arterial stresses were observed in the artery supported by the welded stent. The peak stress concentration may increase the occurrence of neointimal hyperplasia.

AB - Braided wire stents demonstrate distinct characteristics compared to welded ones. In this study, both braided and welded wire stents with the same nominal dimensions were crimped inside a sheath and then deployed into a stenosed artery using finite element analysis. The braided wire stent was generated by overlapping wires to form crisscross shape. A welded wire stent was created by welding the intersection points of wires to avoid sliding between wires. The effect of fabrication technique on mechanical behavior of Nitinol wire stents was evaluated. The results showed that relative sliding between wires reduced the deformation of the braided stent, which led to less radial strength than the welded one; therefore, the deployed braided stent was more conformed to the anatomic shape of the lesion and much less efficient for restoring the patency of the stenotic artery. Post balloon-dilation was commonly used to improve its performance in terms of lumen gain and deployed shape of the stent. On the contrary, the welded wire stent exhibited a high capacity for pushing the occlusion outward. It reached an approximately uniform shape after deployment. The welded joints caused larger deformation and high strain on the stent struts, which indicate a potential earlier failure for the welded stent. In addition, higher contact pressure at the stent-lesion interface and higher arterial stresses were observed in the artery supported by the welded stent. The peak stress concentration may increase the occurrence of neointimal hyperplasia.

KW - Braiding

KW - Finite element method

KW - Nitinol stent

KW - Post balloon-dilation

KW - Strain energy

KW - Stress analysis

KW - Welding

UR - http://www.scopus.com/inward/record.url?scp=84859571557&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84859571557&partnerID=8YFLogxK

U2 - 10.1115/1.4005788

DO - 10.1115/1.4005788

M3 - Article

VL - 6

JO - Journal of Medical Devices, Transactions of the ASME

JF - Journal of Medical Devices, Transactions of the ASME

SN - 1932-6181

IS - 1

M1 - 011007

ER -