Testing of the vocalog vocal monitor

Jeff Searl, Angela Dietsch

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Objective To elucidate capabilities and limitations of the VocaLog, a device marketed to log-calibrated dB sound pressure level (SPL). Study Design The study design varied depending on the experiment. All were prospective. Some were case series, and others were cohort studies without controls. Method Experiments were conducted to determine (1) whether the VocaLog logged phonatory activity and silence when it should, (2) if nonphonatory activities were detected, (3) correlation of VocaLog dB values to an external sound level meter (SLM), and (4) accuracy of phonation time (PT) and speaking time (ST) estimates from the VocaLog. Results Silence and phonatory activity were logged as such nearly 100% of the time. Nonphonatory activities were sometimes detected as dB values, including coughs, throat clear, belching, and swallows. The dB values from the VocaLog were strongly correlated with dB SPL from an external SLM. When on the neck, the device rarely picked up external sounds when the external noise was between 85 and 103 dB SPL. The VocaLog gave a reasonable estimate of ST but overestimated PT. Conclusions Overall, the VocaLog holds promise as means of indexing vocal loudness via calibrated dB SPL levels. However, some nonphonatory activity is also likely to be logged. The device provides a reasonable estimate of ST, but not PT.

Original languageEnglish (US)
Pages (from-to)523.e27-523.e37
JournalJournal of Voice
Volume28
Issue number4
DOIs
StatePublished - Jul 2014

Fingerprint

Phonation
Pressure
Equipment and Supplies
Eructation
Swallows
Pharynx
Cough
Noise
Cohort Studies
Neck

Keywords

  • Instrumentation
  • Sound pressure level
  • Voice monitor

ASJC Scopus subject areas

  • Otorhinolaryngology
  • Speech and Hearing
  • LPN and LVN

Cite this

Testing of the vocalog vocal monitor. / Searl, Jeff; Dietsch, Angela.

In: Journal of Voice, Vol. 28, No. 4, 07.2014, p. 523.e27-523.e37.

Research output: Contribution to journalArticle

Searl, Jeff ; Dietsch, Angela. / Testing of the vocalog vocal monitor. In: Journal of Voice. 2014 ; Vol. 28, No. 4. pp. 523.e27-523.e37.
@article{7a9e8c71b7b647e3bbc23beefefc9acd,
title = "Testing of the vocalog vocal monitor",
abstract = "Objective To elucidate capabilities and limitations of the VocaLog, a device marketed to log-calibrated dB sound pressure level (SPL). Study Design The study design varied depending on the experiment. All were prospective. Some were case series, and others were cohort studies without controls. Method Experiments were conducted to determine (1) whether the VocaLog logged phonatory activity and silence when it should, (2) if nonphonatory activities were detected, (3) correlation of VocaLog dB values to an external sound level meter (SLM), and (4) accuracy of phonation time (PT) and speaking time (ST) estimates from the VocaLog. Results Silence and phonatory activity were logged as such nearly 100{\%} of the time. Nonphonatory activities were sometimes detected as dB values, including coughs, throat clear, belching, and swallows. The dB values from the VocaLog were strongly correlated with dB SPL from an external SLM. When on the neck, the device rarely picked up external sounds when the external noise was between 85 and 103 dB SPL. The VocaLog gave a reasonable estimate of ST but overestimated PT. Conclusions Overall, the VocaLog holds promise as means of indexing vocal loudness via calibrated dB SPL levels. However, some nonphonatory activity is also likely to be logged. The device provides a reasonable estimate of ST, but not PT.",
keywords = "Instrumentation, Sound pressure level, Voice monitor",
author = "Jeff Searl and Angela Dietsch",
year = "2014",
month = "7",
doi = "10.1016/j.jvoice.2014.01.009",
language = "English (US)",
volume = "28",
pages = "523.e27--523.e37",
journal = "Journal of Voice",
issn = "0892-1997",
publisher = "Mosby Inc.",
number = "4",

}

TY - JOUR

T1 - Testing of the vocalog vocal monitor

AU - Searl, Jeff

AU - Dietsch, Angela

PY - 2014/7

Y1 - 2014/7

N2 - Objective To elucidate capabilities and limitations of the VocaLog, a device marketed to log-calibrated dB sound pressure level (SPL). Study Design The study design varied depending on the experiment. All were prospective. Some were case series, and others were cohort studies without controls. Method Experiments were conducted to determine (1) whether the VocaLog logged phonatory activity and silence when it should, (2) if nonphonatory activities were detected, (3) correlation of VocaLog dB values to an external sound level meter (SLM), and (4) accuracy of phonation time (PT) and speaking time (ST) estimates from the VocaLog. Results Silence and phonatory activity were logged as such nearly 100% of the time. Nonphonatory activities were sometimes detected as dB values, including coughs, throat clear, belching, and swallows. The dB values from the VocaLog were strongly correlated with dB SPL from an external SLM. When on the neck, the device rarely picked up external sounds when the external noise was between 85 and 103 dB SPL. The VocaLog gave a reasonable estimate of ST but overestimated PT. Conclusions Overall, the VocaLog holds promise as means of indexing vocal loudness via calibrated dB SPL levels. However, some nonphonatory activity is also likely to be logged. The device provides a reasonable estimate of ST, but not PT.

AB - Objective To elucidate capabilities and limitations of the VocaLog, a device marketed to log-calibrated dB sound pressure level (SPL). Study Design The study design varied depending on the experiment. All were prospective. Some were case series, and others were cohort studies without controls. Method Experiments were conducted to determine (1) whether the VocaLog logged phonatory activity and silence when it should, (2) if nonphonatory activities were detected, (3) correlation of VocaLog dB values to an external sound level meter (SLM), and (4) accuracy of phonation time (PT) and speaking time (ST) estimates from the VocaLog. Results Silence and phonatory activity were logged as such nearly 100% of the time. Nonphonatory activities were sometimes detected as dB values, including coughs, throat clear, belching, and swallows. The dB values from the VocaLog were strongly correlated with dB SPL from an external SLM. When on the neck, the device rarely picked up external sounds when the external noise was between 85 and 103 dB SPL. The VocaLog gave a reasonable estimate of ST but overestimated PT. Conclusions Overall, the VocaLog holds promise as means of indexing vocal loudness via calibrated dB SPL levels. However, some nonphonatory activity is also likely to be logged. The device provides a reasonable estimate of ST, but not PT.

KW - Instrumentation

KW - Sound pressure level

KW - Voice monitor

UR - http://www.scopus.com/inward/record.url?scp=84902546116&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84902546116&partnerID=8YFLogxK

U2 - 10.1016/j.jvoice.2014.01.009

DO - 10.1016/j.jvoice.2014.01.009

M3 - Article

C2 - 24836363

AN - SCOPUS:84902546116

VL - 28

SP - 523.e27-523.e37

JO - Journal of Voice

JF - Journal of Voice

SN - 0892-1997

IS - 4

ER -