SU‐GG‐T‐553: Dosimetry Impact of Radiation Beams Due to Presence of Calypso Electromagnetic Array

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

Purpose: The 4D Calypso systems are used for accurately localizing and continuously tracking the targets in prostate treatments. An electromagnetic array composed of low density materials is positioned over the patient during radiation treatment delivery to excite the Beacon transponders, which are permanently implanted in targets, and sense their motions. This study is to investigate and verify the attenuation of photon beams due to presence of the array. Method and Materials: A solid water phantom was used to simulate the radiation treatment. The array was positioned at distant of 18cm to isocenter and aligned to ceiling laser. The 0.3cc cylinder chamber was used for measurement of the attenuation for field size of 5×5cm2 and 10×10cm2 and 0.015cc pinpoint chamber for field size of 2×2 cm2 and IMRT fields. For each field, chambers were at depth of 8cm, 10cm, and 12cm respectively and the 6MV beams delivered 200MUs at gantry angles of 0, 10, 20, 30, and 40 degree. Each measurement was repeated twice. The attenuation was evaluated by comparison of dose measured at isocenter with and without the array. Results: The measurements show that the attenuation from beams normal to the array ranges from 1% to 2%. This result is consistent to conclusion provided in Calypso user manual (<2% at 6 cm depth), but disagrees with another published results (<0.5%, Med Phys, 30(6), 1473 (2003)). The results demonstrate that attenuations increase with gantry angle and are larger than 2% at 40 degree for field size of 10×10cm2. Measurements with IMRT beams show that the attenuations are less than 2% at gantry angle less that 20 degree and could be as high as 3% at 40 degree. Conclusion: To minimize the array impact to radiation delivery, the large oblique angle should be avoided for the beams incident to the array.

Original languageEnglish (US)
Number of pages1
JournalMedical physics
Volume35
Issue number6
DOIs
StatePublished - Jun 2008

Fingerprint

Radiometry
Electromagnetic Phenomena
Radiation
Photons
Prostate
Lasers
Therapeutics
Water
thiacloprid

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

@article{8fcedfd551a24f8a86ded29a9f9f6467,
title = "SU‐GG‐T‐553: Dosimetry Impact of Radiation Beams Due to Presence of Calypso Electromagnetic Array",
abstract = "Purpose: The 4D Calypso systems are used for accurately localizing and continuously tracking the targets in prostate treatments. An electromagnetic array composed of low density materials is positioned over the patient during radiation treatment delivery to excite the Beacon transponders, which are permanently implanted in targets, and sense their motions. This study is to investigate and verify the attenuation of photon beams due to presence of the array. Method and Materials: A solid water phantom was used to simulate the radiation treatment. The array was positioned at distant of 18cm to isocenter and aligned to ceiling laser. The 0.3cc cylinder chamber was used for measurement of the attenuation for field size of 5×5cm2 and 10×10cm2 and 0.015cc pinpoint chamber for field size of 2×2 cm2 and IMRT fields. For each field, chambers were at depth of 8cm, 10cm, and 12cm respectively and the 6MV beams delivered 200MUs at gantry angles of 0, 10, 20, 30, and 40 degree. Each measurement was repeated twice. The attenuation was evaluated by comparison of dose measured at isocenter with and without the array. Results: The measurements show that the attenuation from beams normal to the array ranges from 1{\%} to 2{\%}. This result is consistent to conclusion provided in Calypso user manual (<2{\%} at 6 cm depth), but disagrees with another published results (<0.5{\%}, Med Phys, 30(6), 1473 (2003)). The results demonstrate that attenuations increase with gantry angle and are larger than 2{\%} at 40 degree for field size of 10×10cm2. Measurements with IMRT beams show that the attenuations are less than 2{\%} at gantry angle less that 20 degree and could be as high as 3{\%} at 40 degree. Conclusion: To minimize the array impact to radiation delivery, the large oblique angle should be avoided for the beams incident to the array.",
author = "Sicong Li and Enke, {Charles Arthur}",
year = "2008",
month = "6",
doi = "10.1118/1.2962302",
language = "English (US)",
volume = "35",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐GG‐T‐553

T2 - Dosimetry Impact of Radiation Beams Due to Presence of Calypso Electromagnetic Array

AU - Li, Sicong

AU - Enke, Charles Arthur

PY - 2008/6

Y1 - 2008/6

N2 - Purpose: The 4D Calypso systems are used for accurately localizing and continuously tracking the targets in prostate treatments. An electromagnetic array composed of low density materials is positioned over the patient during radiation treatment delivery to excite the Beacon transponders, which are permanently implanted in targets, and sense their motions. This study is to investigate and verify the attenuation of photon beams due to presence of the array. Method and Materials: A solid water phantom was used to simulate the radiation treatment. The array was positioned at distant of 18cm to isocenter and aligned to ceiling laser. The 0.3cc cylinder chamber was used for measurement of the attenuation for field size of 5×5cm2 and 10×10cm2 and 0.015cc pinpoint chamber for field size of 2×2 cm2 and IMRT fields. For each field, chambers were at depth of 8cm, 10cm, and 12cm respectively and the 6MV beams delivered 200MUs at gantry angles of 0, 10, 20, 30, and 40 degree. Each measurement was repeated twice. The attenuation was evaluated by comparison of dose measured at isocenter with and without the array. Results: The measurements show that the attenuation from beams normal to the array ranges from 1% to 2%. This result is consistent to conclusion provided in Calypso user manual (<2% at 6 cm depth), but disagrees with another published results (<0.5%, Med Phys, 30(6), 1473 (2003)). The results demonstrate that attenuations increase with gantry angle and are larger than 2% at 40 degree for field size of 10×10cm2. Measurements with IMRT beams show that the attenuations are less than 2% at gantry angle less that 20 degree and could be as high as 3% at 40 degree. Conclusion: To minimize the array impact to radiation delivery, the large oblique angle should be avoided for the beams incident to the array.

AB - Purpose: The 4D Calypso systems are used for accurately localizing and continuously tracking the targets in prostate treatments. An electromagnetic array composed of low density materials is positioned over the patient during radiation treatment delivery to excite the Beacon transponders, which are permanently implanted in targets, and sense their motions. This study is to investigate and verify the attenuation of photon beams due to presence of the array. Method and Materials: A solid water phantom was used to simulate the radiation treatment. The array was positioned at distant of 18cm to isocenter and aligned to ceiling laser. The 0.3cc cylinder chamber was used for measurement of the attenuation for field size of 5×5cm2 and 10×10cm2 and 0.015cc pinpoint chamber for field size of 2×2 cm2 and IMRT fields. For each field, chambers were at depth of 8cm, 10cm, and 12cm respectively and the 6MV beams delivered 200MUs at gantry angles of 0, 10, 20, 30, and 40 degree. Each measurement was repeated twice. The attenuation was evaluated by comparison of dose measured at isocenter with and without the array. Results: The measurements show that the attenuation from beams normal to the array ranges from 1% to 2%. This result is consistent to conclusion provided in Calypso user manual (<2% at 6 cm depth), but disagrees with another published results (<0.5%, Med Phys, 30(6), 1473 (2003)). The results demonstrate that attenuations increase with gantry angle and are larger than 2% at 40 degree for field size of 10×10cm2. Measurements with IMRT beams show that the attenuations are less than 2% at gantry angle less that 20 degree and could be as high as 3% at 40 degree. Conclusion: To minimize the array impact to radiation delivery, the large oblique angle should be avoided for the beams incident to the array.

UR - http://www.scopus.com/inward/record.url?scp=84902190422&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84902190422&partnerID=8YFLogxK

U2 - 10.1118/1.2962302

DO - 10.1118/1.2962302

M3 - Article

AN - SCOPUS:84902190422

VL - 35

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -