Study of scan protocol for exposure reduction in hybrid spectral micro-CT

Alex M.T. Opie, James R. Bennett, Michael Walsh, Kishore Rajendran, Hengyong Yu, Qiong Xu, Anthony Butler, Philip Butler, Guohua Cao, Aaron M Mohs, Ge Wang

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The hybrid spectral micro-computed tomography (CT) architecture integrates a conventional imaging chain and an interior spectral imaging chain, and has been proven to be an important development in spectral CT. The motivation for this study is to minimize X-ray exposure for hybrid spectral micro-CT using both simulated and experimental scan data while maintaining the spectral fidelity of the reconstruction. Three elements of the hybrid scan protocol are investigated: truncation of the interior spectral chain and the numbers of projections for each of the global and interior imaging chains. The effect of these elements is quantified by analyzing how each affects the reconstructed spectral accuracy. The results demonstrate that there is significant scope for reduction of radiation exposure in the hybrid scan protocol. It appears decreasing the number of conventional projections offers the most potential for exposure reduction, while further reduction is possible by decreasing the interior FOV and number of spectral projections.

Original languageEnglish (US)
Pages (from-to)444-455
Number of pages12
JournalScanning
Volume36
Issue number4
DOIs
StatePublished - Jan 1 2014

Fingerprint

Tomography
tomography
Imaging techniques
projection
radiation dosage
Radiation
X rays
X-Rays
approximation
x rays

Keywords

  • image reconstruction
  • micro-computed tomography
  • radiation exposure
  • spectral imaging

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • Instrumentation

Cite this

Opie, A. M. T., Bennett, J. R., Walsh, M., Rajendran, K., Yu, H., Xu, Q., ... Wang, G. (2014). Study of scan protocol for exposure reduction in hybrid spectral micro-CT. Scanning, 36(4), 444-455. https://doi.org/10.1002/sca.21140

Study of scan protocol for exposure reduction in hybrid spectral micro-CT. / Opie, Alex M.T.; Bennett, James R.; Walsh, Michael; Rajendran, Kishore; Yu, Hengyong; Xu, Qiong; Butler, Anthony; Butler, Philip; Cao, Guohua; Mohs, Aaron M; Wang, Ge.

In: Scanning, Vol. 36, No. 4, 01.01.2014, p. 444-455.

Research output: Contribution to journalArticle

Opie, AMT, Bennett, JR, Walsh, M, Rajendran, K, Yu, H, Xu, Q, Butler, A, Butler, P, Cao, G, Mohs, AM & Wang, G 2014, 'Study of scan protocol for exposure reduction in hybrid spectral micro-CT', Scanning, vol. 36, no. 4, pp. 444-455. https://doi.org/10.1002/sca.21140
Opie AMT, Bennett JR, Walsh M, Rajendran K, Yu H, Xu Q et al. Study of scan protocol for exposure reduction in hybrid spectral micro-CT. Scanning. 2014 Jan 1;36(4):444-455. https://doi.org/10.1002/sca.21140
Opie, Alex M.T. ; Bennett, James R. ; Walsh, Michael ; Rajendran, Kishore ; Yu, Hengyong ; Xu, Qiong ; Butler, Anthony ; Butler, Philip ; Cao, Guohua ; Mohs, Aaron M ; Wang, Ge. / Study of scan protocol for exposure reduction in hybrid spectral micro-CT. In: Scanning. 2014 ; Vol. 36, No. 4. pp. 444-455.
@article{8f7d76dfb7a94fd592eeb9fda08c1103,
title = "Study of scan protocol for exposure reduction in hybrid spectral micro-CT",
abstract = "The hybrid spectral micro-computed tomography (CT) architecture integrates a conventional imaging chain and an interior spectral imaging chain, and has been proven to be an important development in spectral CT. The motivation for this study is to minimize X-ray exposure for hybrid spectral micro-CT using both simulated and experimental scan data while maintaining the spectral fidelity of the reconstruction. Three elements of the hybrid scan protocol are investigated: truncation of the interior spectral chain and the numbers of projections for each of the global and interior imaging chains. The effect of these elements is quantified by analyzing how each affects the reconstructed spectral accuracy. The results demonstrate that there is significant scope for reduction of radiation exposure in the hybrid scan protocol. It appears decreasing the number of conventional projections offers the most potential for exposure reduction, while further reduction is possible by decreasing the interior FOV and number of spectral projections.",
keywords = "image reconstruction, micro-computed tomography, radiation exposure, spectral imaging",
author = "Opie, {Alex M.T.} and Bennett, {James R.} and Michael Walsh and Kishore Rajendran and Hengyong Yu and Qiong Xu and Anthony Butler and Philip Butler and Guohua Cao and Mohs, {Aaron M} and Ge Wang",
year = "2014",
month = "1",
day = "1",
doi = "10.1002/sca.21140",
language = "English (US)",
volume = "36",
pages = "444--455",
journal = "Scanning",
issn = "0161-0457",
publisher = "John Wiley and Sons Inc.",
number = "4",

}

TY - JOUR

T1 - Study of scan protocol for exposure reduction in hybrid spectral micro-CT

AU - Opie, Alex M.T.

AU - Bennett, James R.

AU - Walsh, Michael

AU - Rajendran, Kishore

AU - Yu, Hengyong

AU - Xu, Qiong

AU - Butler, Anthony

AU - Butler, Philip

AU - Cao, Guohua

AU - Mohs, Aaron M

AU - Wang, Ge

PY - 2014/1/1

Y1 - 2014/1/1

N2 - The hybrid spectral micro-computed tomography (CT) architecture integrates a conventional imaging chain and an interior spectral imaging chain, and has been proven to be an important development in spectral CT. The motivation for this study is to minimize X-ray exposure for hybrid spectral micro-CT using both simulated and experimental scan data while maintaining the spectral fidelity of the reconstruction. Three elements of the hybrid scan protocol are investigated: truncation of the interior spectral chain and the numbers of projections for each of the global and interior imaging chains. The effect of these elements is quantified by analyzing how each affects the reconstructed spectral accuracy. The results demonstrate that there is significant scope for reduction of radiation exposure in the hybrid scan protocol. It appears decreasing the number of conventional projections offers the most potential for exposure reduction, while further reduction is possible by decreasing the interior FOV and number of spectral projections.

AB - The hybrid spectral micro-computed tomography (CT) architecture integrates a conventional imaging chain and an interior spectral imaging chain, and has been proven to be an important development in spectral CT. The motivation for this study is to minimize X-ray exposure for hybrid spectral micro-CT using both simulated and experimental scan data while maintaining the spectral fidelity of the reconstruction. Three elements of the hybrid scan protocol are investigated: truncation of the interior spectral chain and the numbers of projections for each of the global and interior imaging chains. The effect of these elements is quantified by analyzing how each affects the reconstructed spectral accuracy. The results demonstrate that there is significant scope for reduction of radiation exposure in the hybrid scan protocol. It appears decreasing the number of conventional projections offers the most potential for exposure reduction, while further reduction is possible by decreasing the interior FOV and number of spectral projections.

KW - image reconstruction

KW - micro-computed tomography

KW - radiation exposure

KW - spectral imaging

UR - http://www.scopus.com/inward/record.url?scp=84905719948&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84905719948&partnerID=8YFLogxK

U2 - 10.1002/sca.21140

DO - 10.1002/sca.21140

M3 - Article

VL - 36

SP - 444

EP - 455

JO - Scanning

JF - Scanning

SN - 0161-0457

IS - 4

ER -