Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM)

Shih Ying Huang, John M. Boone, Dandan Zheng, Kai Yang, Nathan J. Packard, George Burkett

Research output: Chapter in Book/Report/Conference proceedingConference contribution

3 Citations (Scopus)

Abstract

Contrast-enhanced dual energy mammography (CEDEM) has been shown effective in enhancing the visibility of breast cancer in diagnostic mammographic images. The simulation and phantom studies of CEDEM were implemented to enhance acquisition techniques of a dedicated cone-beam breast CT scanner fabricated in our laboratory. CEDEM images were computer-simulated numerically, and the results were compared and validated with physical experiments using a custom-made breast-equivalent phantom with pseudo anatomical structures. The technique factors to be optimized included low- and high-energy x-ray spectra (kVp), spectra filters and mAs (x-ray tube current × scan time in second). Signal-difference-to-noise ratio (SDNR) of the log-weighted subtracted image was computed to optimize subtraction parameters using several combinations of breast compositions and thicknesses. Both simulation and phantom studies suggest that 40 kVp (filtered with 1.5 mm Al) and 50 kVp (filtered with 1.5 mm Al and 0.3 mm Cu) were the optimal spectra with the highest SDNR on the dual energy subtracted image of a 6-cm, 50% glandular breast. Breast composition and thickness were found to have little influence on the optimized technique factors. We conclude that the simulation and phantom studies demonstrated the effectiveness of CEDEM using the prototype breast CT scanner in producing high-quality, lesion-enhanced images.

Original languageEnglish (US)
Title of host publicationDigital Mammography - 9th International Workshop, IWDM 2008, Proceedings
Pages124-129
Number of pages6
DOIs
StatePublished - Sep 9 2008
Event9th International Workshop on Digital Mammography, IWDM 2008 - Tucson, AZ, United States
Duration: Jul 20 2008Jul 23 2008

Publication series

NameLecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)
Volume5116 LNCS
ISSN (Print)0302-9743
ISSN (Electronic)1611-3349

Conference

Conference9th International Workshop on Digital Mammography, IWDM 2008
CountryUnited States
CityTucson, AZ
Period7/20/087/23/08

Fingerprint

Mammography
Phantom
Breast
Energy
Simulation
Signal-To-Noise Ratio
Scanner
X rays
Wave filters
Chemical analysis
X-Rays
Visibility
Cones
Cone-Beam Computed Tomography
Subtraction
Breast Cancer
High Energy
Diagnostics
Tube
Cone

Keywords

  • Anatomical breast phantom
  • Computer simula-tion
  • Contrast-enhanced dual energy mammography

ASJC Scopus subject areas

  • Theoretical Computer Science
  • Computer Science(all)

Cite this

Huang, S. Y., Boone, J. M., Zheng, D., Yang, K., Packard, N. J., & Burkett, G. (2008). Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM). In Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings (pp. 124-129). (Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics); Vol. 5116 LNCS). https://doi.org/10.1007/978-3-540-70538-3_18

Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM). / Huang, Shih Ying; Boone, John M.; Zheng, Dandan; Yang, Kai; Packard, Nathan J.; Burkett, George.

Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings. 2008. p. 124-129 (Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics); Vol. 5116 LNCS).

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Huang, SY, Boone, JM, Zheng, D, Yang, K, Packard, NJ & Burkett, G 2008, Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM). in Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings. Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics), vol. 5116 LNCS, pp. 124-129, 9th International Workshop on Digital Mammography, IWDM 2008, Tucson, AZ, United States, 7/20/08. https://doi.org/10.1007/978-3-540-70538-3_18
Huang SY, Boone JM, Zheng D, Yang K, Packard NJ, Burkett G. Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM). In Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings. 2008. p. 124-129. (Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)). https://doi.org/10.1007/978-3-540-70538-3_18
Huang, Shih Ying ; Boone, John M. ; Zheng, Dandan ; Yang, Kai ; Packard, Nathan J. ; Burkett, George. / Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM). Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings. 2008. pp. 124-129 (Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)).
@inproceedings{01b54a5f19834d4c93f5c3e290491b66,
title = "Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM)",
abstract = "Contrast-enhanced dual energy mammography (CEDEM) has been shown effective in enhancing the visibility of breast cancer in diagnostic mammographic images. The simulation and phantom studies of CEDEM were implemented to enhance acquisition techniques of a dedicated cone-beam breast CT scanner fabricated in our laboratory. CEDEM images were computer-simulated numerically, and the results were compared and validated with physical experiments using a custom-made breast-equivalent phantom with pseudo anatomical structures. The technique factors to be optimized included low- and high-energy x-ray spectra (kVp), spectra filters and mAs (x-ray tube current × scan time in second). Signal-difference-to-noise ratio (SDNR) of the log-weighted subtracted image was computed to optimize subtraction parameters using several combinations of breast compositions and thicknesses. Both simulation and phantom studies suggest that 40 kVp (filtered with 1.5 mm Al) and 50 kVp (filtered with 1.5 mm Al and 0.3 mm Cu) were the optimal spectra with the highest SDNR on the dual energy subtracted image of a 6-cm, 50{\%} glandular breast. Breast composition and thickness were found to have little influence on the optimized technique factors. We conclude that the simulation and phantom studies demonstrated the effectiveness of CEDEM using the prototype breast CT scanner in producing high-quality, lesion-enhanced images.",
keywords = "Anatomical breast phantom, Computer simula-tion, Contrast-enhanced dual energy mammography",
author = "Huang, {Shih Ying} and Boone, {John M.} and Dandan Zheng and Kai Yang and Packard, {Nathan J.} and George Burkett",
year = "2008",
month = "9",
day = "9",
doi = "10.1007/978-3-540-70538-3_18",
language = "English (US)",
isbn = "3540705376",
series = "Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)",
pages = "124--129",
booktitle = "Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings",

}

TY - GEN

T1 - Simulation and phantom studies of Contrast-Enhanced Dual Energy Mammography (CEDEM)

AU - Huang, Shih Ying

AU - Boone, John M.

AU - Zheng, Dandan

AU - Yang, Kai

AU - Packard, Nathan J.

AU - Burkett, George

PY - 2008/9/9

Y1 - 2008/9/9

N2 - Contrast-enhanced dual energy mammography (CEDEM) has been shown effective in enhancing the visibility of breast cancer in diagnostic mammographic images. The simulation and phantom studies of CEDEM were implemented to enhance acquisition techniques of a dedicated cone-beam breast CT scanner fabricated in our laboratory. CEDEM images were computer-simulated numerically, and the results were compared and validated with physical experiments using a custom-made breast-equivalent phantom with pseudo anatomical structures. The technique factors to be optimized included low- and high-energy x-ray spectra (kVp), spectra filters and mAs (x-ray tube current × scan time in second). Signal-difference-to-noise ratio (SDNR) of the log-weighted subtracted image was computed to optimize subtraction parameters using several combinations of breast compositions and thicknesses. Both simulation and phantom studies suggest that 40 kVp (filtered with 1.5 mm Al) and 50 kVp (filtered with 1.5 mm Al and 0.3 mm Cu) were the optimal spectra with the highest SDNR on the dual energy subtracted image of a 6-cm, 50% glandular breast. Breast composition and thickness were found to have little influence on the optimized technique factors. We conclude that the simulation and phantom studies demonstrated the effectiveness of CEDEM using the prototype breast CT scanner in producing high-quality, lesion-enhanced images.

AB - Contrast-enhanced dual energy mammography (CEDEM) has been shown effective in enhancing the visibility of breast cancer in diagnostic mammographic images. The simulation and phantom studies of CEDEM were implemented to enhance acquisition techniques of a dedicated cone-beam breast CT scanner fabricated in our laboratory. CEDEM images were computer-simulated numerically, and the results were compared and validated with physical experiments using a custom-made breast-equivalent phantom with pseudo anatomical structures. The technique factors to be optimized included low- and high-energy x-ray spectra (kVp), spectra filters and mAs (x-ray tube current × scan time in second). Signal-difference-to-noise ratio (SDNR) of the log-weighted subtracted image was computed to optimize subtraction parameters using several combinations of breast compositions and thicknesses. Both simulation and phantom studies suggest that 40 kVp (filtered with 1.5 mm Al) and 50 kVp (filtered with 1.5 mm Al and 0.3 mm Cu) were the optimal spectra with the highest SDNR on the dual energy subtracted image of a 6-cm, 50% glandular breast. Breast composition and thickness were found to have little influence on the optimized technique factors. We conclude that the simulation and phantom studies demonstrated the effectiveness of CEDEM using the prototype breast CT scanner in producing high-quality, lesion-enhanced images.

KW - Anatomical breast phantom

KW - Computer simula-tion

KW - Contrast-enhanced dual energy mammography

UR - http://www.scopus.com/inward/record.url?scp=50949098877&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=50949098877&partnerID=8YFLogxK

U2 - 10.1007/978-3-540-70538-3_18

DO - 10.1007/978-3-540-70538-3_18

M3 - Conference contribution

AN - SCOPUS:50949098877

SN - 3540705376

SN - 9783540705376

T3 - Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics)

SP - 124

EP - 129

BT - Digital Mammography - 9th International Workshop, IWDM 2008, Proceedings

ER -