The electromagnetic field focusing (EFF) probe, based on the principle of eddy current convergence, produces pinpoint heat at its point of contact so intense that it can thrombose vessels. Since the probe diameter can be as small as 0.2 mm and can be flexible, it can be navigated transvascularly to a desired point in order to thrombose vessels. The present experiments were conducted to study intravascular thrombosis of arteries. Six arteries in 2 dogs were used for the study. The tip of a flexible probe with a probe tip length of 7 mm was placed in the artery at a desired point and exposed to 150 watts of electromagnetic field. Arteries that were 3 millimeters (mm) in diameter could be thrombosed when exposed to the electromagnetic field for twenty to thirty seconds (4 arter ies). Fifty seconds of exposure resulted in perforation of the artery (1 artery). A larger artery (6 mm in diameter) could not be thrombosed with this method. Histologic studies showed a firm thrombus with no changes in the arterial wall. This technique could be used to thrombose arteries feeding vascular tumors, arteriovenous malformations, and certain intracranial aneurysms, and for arte rial embolization in bleeding problems.

Original languageEnglish (US)
Pages (from-to)245-250
Number of pages6
JournalVascular and Endovascular Surgery
Volume24
Issue number4
DOIs
StatePublished - Jan 1 1990

Fingerprint

Electromagnetic Fields
Feasibility Studies
Thrombosis
Arteries
Intracranial Arteriovenous Malformations
Intracranial Aneurysm
Blood Vessels
Hot Temperature
Dogs
Hemorrhage

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Surgery

Cite this

Selective Arterial Thrombosis Using the Electromagnetic Field Focusing (EFF) System : A Feasibility Study. / Patil, Arun-Angelo; Hahn, Francis; Yamanashi, W.

In: Vascular and Endovascular Surgery, Vol. 24, No. 4, 01.01.1990, p. 245-250.

Research output: Contribution to journalArticle

Patil, Arun-Angelo ; Hahn, Francis ; Yamanashi, W. / Selective Arterial Thrombosis Using the Electromagnetic Field Focusing (EFF) System : A Feasibility Study. In: Vascular and Endovascular Surgery. 1990 ; Vol. 24, No. 4. pp. 245-250.
@article{27f9bd0575654ee59a8c9d85ef055cbe,
title = "Selective Arterial Thrombosis Using the Electromagnetic Field Focusing (EFF) System: A Feasibility Study",
abstract = "The electromagnetic field focusing (EFF) probe, based on the principle of eddy current convergence, produces pinpoint heat at its point of contact so intense that it can thrombose vessels. Since the probe diameter can be as small as 0.2 mm and can be flexible, it can be navigated transvascularly to a desired point in order to thrombose vessels. The present experiments were conducted to study intravascular thrombosis of arteries. Six arteries in 2 dogs were used for the study. The tip of a flexible probe with a probe tip length of 7 mm was placed in the artery at a desired point and exposed to 150 watts of electromagnetic field. Arteries that were 3 millimeters (mm) in diameter could be thrombosed when exposed to the electromagnetic field for twenty to thirty seconds (4 arter ies). Fifty seconds of exposure resulted in perforation of the artery (1 artery). A larger artery (6 mm in diameter) could not be thrombosed with this method. Histologic studies showed a firm thrombus with no changes in the arterial wall. This technique could be used to thrombose arteries feeding vascular tumors, arteriovenous malformations, and certain intracranial aneurysms, and for arte rial embolization in bleeding problems.",
author = "Arun-Angelo Patil and Francis Hahn and W. Yamanashi",
year = "1990",
month = "1",
day = "1",
doi = "10.1177/153857449002400405",
language = "English (US)",
volume = "24",
pages = "245--250",
journal = "Vascular and Endovascular Surgery",
issn = "1538-5744",
publisher = "SAGE Publications Inc.",
number = "4",

}

TY - JOUR

T1 - Selective Arterial Thrombosis Using the Electromagnetic Field Focusing (EFF) System

T2 - A Feasibility Study

AU - Patil, Arun-Angelo

AU - Hahn, Francis

AU - Yamanashi, W.

PY - 1990/1/1

Y1 - 1990/1/1

N2 - The electromagnetic field focusing (EFF) probe, based on the principle of eddy current convergence, produces pinpoint heat at its point of contact so intense that it can thrombose vessels. Since the probe diameter can be as small as 0.2 mm and can be flexible, it can be navigated transvascularly to a desired point in order to thrombose vessels. The present experiments were conducted to study intravascular thrombosis of arteries. Six arteries in 2 dogs were used for the study. The tip of a flexible probe with a probe tip length of 7 mm was placed in the artery at a desired point and exposed to 150 watts of electromagnetic field. Arteries that were 3 millimeters (mm) in diameter could be thrombosed when exposed to the electromagnetic field for twenty to thirty seconds (4 arter ies). Fifty seconds of exposure resulted in perforation of the artery (1 artery). A larger artery (6 mm in diameter) could not be thrombosed with this method. Histologic studies showed a firm thrombus with no changes in the arterial wall. This technique could be used to thrombose arteries feeding vascular tumors, arteriovenous malformations, and certain intracranial aneurysms, and for arte rial embolization in bleeding problems.

AB - The electromagnetic field focusing (EFF) probe, based on the principle of eddy current convergence, produces pinpoint heat at its point of contact so intense that it can thrombose vessels. Since the probe diameter can be as small as 0.2 mm and can be flexible, it can be navigated transvascularly to a desired point in order to thrombose vessels. The present experiments were conducted to study intravascular thrombosis of arteries. Six arteries in 2 dogs were used for the study. The tip of a flexible probe with a probe tip length of 7 mm was placed in the artery at a desired point and exposed to 150 watts of electromagnetic field. Arteries that were 3 millimeters (mm) in diameter could be thrombosed when exposed to the electromagnetic field for twenty to thirty seconds (4 arter ies). Fifty seconds of exposure resulted in perforation of the artery (1 artery). A larger artery (6 mm in diameter) could not be thrombosed with this method. Histologic studies showed a firm thrombus with no changes in the arterial wall. This technique could be used to thrombose arteries feeding vascular tumors, arteriovenous malformations, and certain intracranial aneurysms, and for arte rial embolization in bleeding problems.

UR - http://www.scopus.com/inward/record.url?scp=0025314199&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0025314199&partnerID=8YFLogxK

U2 - 10.1177/153857449002400405

DO - 10.1177/153857449002400405

M3 - Article

VL - 24

SP - 245

EP - 250

JO - Vascular and Endovascular Surgery

JF - Vascular and Endovascular Surgery

SN - 1538-5744

IS - 4

ER -