Relationship between EABR thresholds and levels used to program the CLARION® speech processor

C. J. Brown, S. M. Lopez, Michelle L Hughes, P. J. Abbas

Research output: Contribution to journalArticle

43 Citations (Scopus)

Abstract

The purpose of this study was 1) to describe the relationship between electrically evoked auditory brain stem response (EABR) thresholds and the behavioral measures needed to program the speech processor of the CLARION® Multi-Strategy(TM) Cochlear Implant, and 2) to determine whether the relationship between EABR threshold and the behavioral levels used to program the Clarion device was correlated with the temporal integration abilities of these subjects. The EABR thresholds, psychophysical thresholds, and the Clarion speech processor programming levels were recorded from 29 postlingually deafened adults. The mean EABR thresholds approximated most comfortable levels (M levels) for the programming stimulus. However, intrasubject variability was substantial. Significant intrasubject variability in temporal integration also was observed. Correlations were found between the ability of a cochlear implant user to perform temporal integration and the relationship between his or her EABR threshold and the threshold for the programming stimulus.

Original languageEnglish (US)
Pages (from-to)50-57
Number of pages8
JournalAnnals of Otology, Rhinology and Laryngology
Volume108
Issue number4 II
StatePublished - May 4 1999

Fingerprint

Brain Stem Auditory Evoked Potentials
Aptitude
Cochlear Implants
Equipment and Supplies

Keywords

  • Auditory evoked potential
  • Cochlear implant
  • Electrical stimulation

ASJC Scopus subject areas

  • Otorhinolaryngology

Cite this

Relationship between EABR thresholds and levels used to program the CLARION® speech processor. / Brown, C. J.; Lopez, S. M.; Hughes, Michelle L; Abbas, P. J.

In: Annals of Otology, Rhinology and Laryngology, Vol. 108, No. 4 II, 04.05.1999, p. 50-57.

Research output: Contribution to journalArticle

Brown, C. J. ; Lopez, S. M. ; Hughes, Michelle L ; Abbas, P. J. / Relationship between EABR thresholds and levels used to program the CLARION® speech processor. In: Annals of Otology, Rhinology and Laryngology. 1999 ; Vol. 108, No. 4 II. pp. 50-57.
@article{83674e2883b2456485c54174aac4e473,
title = "Relationship between EABR thresholds and levels used to program the CLARION{\circledR} speech processor",
abstract = "The purpose of this study was 1) to describe the relationship between electrically evoked auditory brain stem response (EABR) thresholds and the behavioral measures needed to program the speech processor of the CLARION{\circledR} Multi-Strategy(TM) Cochlear Implant, and 2) to determine whether the relationship between EABR threshold and the behavioral levels used to program the Clarion device was correlated with the temporal integration abilities of these subjects. The EABR thresholds, psychophysical thresholds, and the Clarion speech processor programming levels were recorded from 29 postlingually deafened adults. The mean EABR thresholds approximated most comfortable levels (M levels) for the programming stimulus. However, intrasubject variability was substantial. Significant intrasubject variability in temporal integration also was observed. Correlations were found between the ability of a cochlear implant user to perform temporal integration and the relationship between his or her EABR threshold and the threshold for the programming stimulus.",
keywords = "Auditory evoked potential, Cochlear implant, Electrical stimulation",
author = "Brown, {C. J.} and Lopez, {S. M.} and Hughes, {Michelle L} and Abbas, {P. J.}",
year = "1999",
month = "5",
day = "4",
language = "English (US)",
volume = "108",
pages = "50--57",
journal = "Annals of Otology, Rhinology and Laryngology",
issn = "0003-4894",
publisher = "Annals Publishing Company",
number = "4 II",

}

TY - JOUR

T1 - Relationship between EABR thresholds and levels used to program the CLARION® speech processor

AU - Brown, C. J.

AU - Lopez, S. M.

AU - Hughes, Michelle L

AU - Abbas, P. J.

PY - 1999/5/4

Y1 - 1999/5/4

N2 - The purpose of this study was 1) to describe the relationship between electrically evoked auditory brain stem response (EABR) thresholds and the behavioral measures needed to program the speech processor of the CLARION® Multi-Strategy(TM) Cochlear Implant, and 2) to determine whether the relationship between EABR threshold and the behavioral levels used to program the Clarion device was correlated with the temporal integration abilities of these subjects. The EABR thresholds, psychophysical thresholds, and the Clarion speech processor programming levels were recorded from 29 postlingually deafened adults. The mean EABR thresholds approximated most comfortable levels (M levels) for the programming stimulus. However, intrasubject variability was substantial. Significant intrasubject variability in temporal integration also was observed. Correlations were found between the ability of a cochlear implant user to perform temporal integration and the relationship between his or her EABR threshold and the threshold for the programming stimulus.

AB - The purpose of this study was 1) to describe the relationship between electrically evoked auditory brain stem response (EABR) thresholds and the behavioral measures needed to program the speech processor of the CLARION® Multi-Strategy(TM) Cochlear Implant, and 2) to determine whether the relationship between EABR threshold and the behavioral levels used to program the Clarion device was correlated with the temporal integration abilities of these subjects. The EABR thresholds, psychophysical thresholds, and the Clarion speech processor programming levels were recorded from 29 postlingually deafened adults. The mean EABR thresholds approximated most comfortable levels (M levels) for the programming stimulus. However, intrasubject variability was substantial. Significant intrasubject variability in temporal integration also was observed. Correlations were found between the ability of a cochlear implant user to perform temporal integration and the relationship between his or her EABR threshold and the threshold for the programming stimulus.

KW - Auditory evoked potential

KW - Cochlear implant

KW - Electrical stimulation

UR - http://www.scopus.com/inward/record.url?scp=0032932785&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0032932785&partnerID=8YFLogxK

M3 - Article

VL - 108

SP - 50

EP - 57

JO - Annals of Otology, Rhinology and Laryngology

JF - Annals of Otology, Rhinology and Laryngology

SN - 0003-4894

IS - 4 II

ER -