Patients receiving parenteral nutrition support frequently undergo operative procedures. Proper selection of patients and close monitoring of glucose levels are crucial if parenteral nutrition is to be maintained intraoperatively. The Dextrometer was used in 10 patients receiving parenteral nutrition intraoperatively. The difference between the Dextrometer readings and the serum glucose level obtained by laboratory analysis was less than 10% in 83% of cases and readings were most accurate in the 200 to 400 mg/dl range. The Dextrometer provides a simple, inexpensive, and reasonably accurate method of monitoring glucose intraoperatively as part of the optimal management of these patients.

Original languageEnglish (US)
Pages (from-to)492-494
Number of pages3
JournalJournal of Parenteral and Enteral Nutrition
Volume7
Issue number5
DOIs
StatePublished - Sep 1983

Fingerprint

Parenteral Nutrition
Glucose
Reading
Operative Surgical Procedures
Physiologic Monitoring
Serum

ASJC Scopus subject areas

  • Medicine (miscellaneous)
  • Nutrition and Dietetics

Cite this

Monitoring Parenteral Nutrition Intraoperatively with the Dextrometer. / Thompson, Jon S; Chapin, James William; Brown, G. L.

In: Journal of Parenteral and Enteral Nutrition, Vol. 7, No. 5, 09.1983, p. 492-494.

Research output: Contribution to journalArticle

@article{2ea231a99e8c46f18fa1129979ea588a,
title = "Monitoring Parenteral Nutrition Intraoperatively with the Dextrometer",
abstract = "Patients receiving parenteral nutrition support frequently undergo operative procedures. Proper selection of patients and close monitoring of glucose levels are crucial if parenteral nutrition is to be maintained intraoperatively. The Dextrometer was used in 10 patients receiving parenteral nutrition intraoperatively. The difference between the Dextrometer readings and the serum glucose level obtained by laboratory analysis was less than 10{\%} in 83{\%} of cases and readings were most accurate in the 200 to 400 mg/dl range. The Dextrometer provides a simple, inexpensive, and reasonably accurate method of monitoring glucose intraoperatively as part of the optimal management of these patients.",
author = "Thompson, {Jon S} and Chapin, {James William} and Brown, {G. L.}",
year = "1983",
month = "9",
doi = "10.1177/0148607183007005492",
language = "English (US)",
volume = "7",
pages = "492--494",
journal = "Journal of Parenteral and Enteral Nutrition",
issn = "0148-6071",
publisher = "SAGE Publications Inc.",
number = "5",

}

TY - JOUR

T1 - Monitoring Parenteral Nutrition Intraoperatively with the Dextrometer

AU - Thompson, Jon S

AU - Chapin, James William

AU - Brown, G. L.

PY - 1983/9

Y1 - 1983/9

N2 - Patients receiving parenteral nutrition support frequently undergo operative procedures. Proper selection of patients and close monitoring of glucose levels are crucial if parenteral nutrition is to be maintained intraoperatively. The Dextrometer was used in 10 patients receiving parenteral nutrition intraoperatively. The difference between the Dextrometer readings and the serum glucose level obtained by laboratory analysis was less than 10% in 83% of cases and readings were most accurate in the 200 to 400 mg/dl range. The Dextrometer provides a simple, inexpensive, and reasonably accurate method of monitoring glucose intraoperatively as part of the optimal management of these patients.

AB - Patients receiving parenteral nutrition support frequently undergo operative procedures. Proper selection of patients and close monitoring of glucose levels are crucial if parenteral nutrition is to be maintained intraoperatively. The Dextrometer was used in 10 patients receiving parenteral nutrition intraoperatively. The difference between the Dextrometer readings and the serum glucose level obtained by laboratory analysis was less than 10% in 83% of cases and readings were most accurate in the 200 to 400 mg/dl range. The Dextrometer provides a simple, inexpensive, and reasonably accurate method of monitoring glucose intraoperatively as part of the optimal management of these patients.

UR - http://www.scopus.com/inward/record.url?scp=0020820172&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020820172&partnerID=8YFLogxK

U2 - 10.1177/0148607183007005492

DO - 10.1177/0148607183007005492

M3 - Article

C2 - 6417367

AN - SCOPUS:0020820172

VL - 7

SP - 492

EP - 494

JO - Journal of Parenteral and Enteral Nutrition

JF - Journal of Parenteral and Enteral Nutrition

SN - 0148-6071

IS - 5

ER -