Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI

Xiaolu Yin, Stephen E. Russek, Gary Zabow, Fan Sun, Jeotikatan Mohapatra, Kathryn E. Keenan, Michael A. Boss, Hao Zeng, J. Ping Liu, Alexandrea Viert, Sy-Hwang Liou, John Moreland

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

Superparamagnetic iron oxide nanoparticles (SPIONs) are widely investigated and utilized as magnetic resonance imaging (MRI) contrast and therapy agents due to their large magnetic moments. Local field inhomogeneities caused by these high magnetic moments are used to generate T2 contrast in clinical high-field MRI, resulting in signal loss (darker contrast). Here we present strong T1 contrast enhancement (brighter contrast) from SPIONs (diameters from 11 nm to 22 nm) as observed in the ultra-low field (ULF) MRI at 0.13 mT. We have achieved a high longitudinal relaxivity for 18 nm SPION solutions, r1 = 615 s−1 mM−1, which is two orders of magnitude larger than typical commercial Gd-based T1 contrast agents operating at high fields (1.5 T and 3 T). The significantly enhanced r1 value at ultra-low fields is attributed to the coupling of proton spins with SPION magnetic fluctuations (Brownian and Néel) associated with a low frequency peak in the imaginary part of AC susceptibility (χ”). SPION-based T1-weighted ULF MRI has the advantages of enhanced signal, shorter imaging times, and iron-oxide-based nontoxic biocompatible agents. This approach shows promise to become a functional imaging technique, similar to PET, where low spatial resolution is compensated for by important functional information.

Original languageEnglish (US)
Article number11863
JournalScientific reports
Volume8
Issue number1
DOIs
StatePublished - Dec 1 2018

Fingerprint

Nanoparticles
Magnetic Resonance Imaging
Contrast Media
ferric oxide
Protons

ASJC Scopus subject areas

  • General

Cite this

Yin, X., Russek, S. E., Zabow, G., Sun, F., Mohapatra, J., Keenan, K. E., ... Moreland, J. (2018). Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI. Scientific reports, 8(1), [11863]. https://doi.org/10.1038/s41598-018-30264-5

Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI. / Yin, Xiaolu; Russek, Stephen E.; Zabow, Gary; Sun, Fan; Mohapatra, Jeotikatan; Keenan, Kathryn E.; Boss, Michael A.; Zeng, Hao; Liu, J. Ping; Viert, Alexandrea; Liou, Sy-Hwang; Moreland, John.

In: Scientific reports, Vol. 8, No. 1, 11863, 01.12.2018.

Research output: Contribution to journalArticle

Yin, X, Russek, SE, Zabow, G, Sun, F, Mohapatra, J, Keenan, KE, Boss, MA, Zeng, H, Liu, JP, Viert, A, Liou, S-H & Moreland, J 2018, 'Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI', Scientific reports, vol. 8, no. 1, 11863. https://doi.org/10.1038/s41598-018-30264-5
Yin, Xiaolu ; Russek, Stephen E. ; Zabow, Gary ; Sun, Fan ; Mohapatra, Jeotikatan ; Keenan, Kathryn E. ; Boss, Michael A. ; Zeng, Hao ; Liu, J. Ping ; Viert, Alexandrea ; Liou, Sy-Hwang ; Moreland, John. / Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI. In: Scientific reports. 2018 ; Vol. 8, No. 1.
@article{a123f077c36449ff993e4912613f3a12,
title = "Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI",
abstract = "Superparamagnetic iron oxide nanoparticles (SPIONs) are widely investigated and utilized as magnetic resonance imaging (MRI) contrast and therapy agents due to their large magnetic moments. Local field inhomogeneities caused by these high magnetic moments are used to generate T2 contrast in clinical high-field MRI, resulting in signal loss (darker contrast). Here we present strong T1 contrast enhancement (brighter contrast) from SPIONs (diameters from 11 nm to 22 nm) as observed in the ultra-low field (ULF) MRI at 0.13 mT. We have achieved a high longitudinal relaxivity for 18 nm SPION solutions, r1 = 615 s−1 mM−1, which is two orders of magnitude larger than typical commercial Gd-based T1 contrast agents operating at high fields (1.5 T and 3 T). The significantly enhanced r1 value at ultra-low fields is attributed to the coupling of proton spins with SPION magnetic fluctuations (Brownian and N{\'e}el) associated with a low frequency peak in the imaginary part of AC susceptibility (χ”). SPION-based T1-weighted ULF MRI has the advantages of enhanced signal, shorter imaging times, and iron-oxide-based nontoxic biocompatible agents. This approach shows promise to become a functional imaging technique, similar to PET, where low spatial resolution is compensated for by important functional information.",
author = "Xiaolu Yin and Russek, {Stephen E.} and Gary Zabow and Fan Sun and Jeotikatan Mohapatra and Keenan, {Kathryn E.} and Boss, {Michael A.} and Hao Zeng and Liu, {J. Ping} and Alexandrea Viert and Sy-Hwang Liou and John Moreland",
year = "2018",
month = "12",
day = "1",
doi = "10.1038/s41598-018-30264-5",
language = "English (US)",
volume = "8",
journal = "Scientific Reports",
issn = "2045-2322",
publisher = "Nature Publishing Group",
number = "1",

}

TY - JOUR

T1 - Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI

AU - Yin, Xiaolu

AU - Russek, Stephen E.

AU - Zabow, Gary

AU - Sun, Fan

AU - Mohapatra, Jeotikatan

AU - Keenan, Kathryn E.

AU - Boss, Michael A.

AU - Zeng, Hao

AU - Liu, J. Ping

AU - Viert, Alexandrea

AU - Liou, Sy-Hwang

AU - Moreland, John

PY - 2018/12/1

Y1 - 2018/12/1

N2 - Superparamagnetic iron oxide nanoparticles (SPIONs) are widely investigated and utilized as magnetic resonance imaging (MRI) contrast and therapy agents due to their large magnetic moments. Local field inhomogeneities caused by these high magnetic moments are used to generate T2 contrast in clinical high-field MRI, resulting in signal loss (darker contrast). Here we present strong T1 contrast enhancement (brighter contrast) from SPIONs (diameters from 11 nm to 22 nm) as observed in the ultra-low field (ULF) MRI at 0.13 mT. We have achieved a high longitudinal relaxivity for 18 nm SPION solutions, r1 = 615 s−1 mM−1, which is two orders of magnitude larger than typical commercial Gd-based T1 contrast agents operating at high fields (1.5 T and 3 T). The significantly enhanced r1 value at ultra-low fields is attributed to the coupling of proton spins with SPION magnetic fluctuations (Brownian and Néel) associated with a low frequency peak in the imaginary part of AC susceptibility (χ”). SPION-based T1-weighted ULF MRI has the advantages of enhanced signal, shorter imaging times, and iron-oxide-based nontoxic biocompatible agents. This approach shows promise to become a functional imaging technique, similar to PET, where low spatial resolution is compensated for by important functional information.

AB - Superparamagnetic iron oxide nanoparticles (SPIONs) are widely investigated and utilized as magnetic resonance imaging (MRI) contrast and therapy agents due to their large magnetic moments. Local field inhomogeneities caused by these high magnetic moments are used to generate T2 contrast in clinical high-field MRI, resulting in signal loss (darker contrast). Here we present strong T1 contrast enhancement (brighter contrast) from SPIONs (diameters from 11 nm to 22 nm) as observed in the ultra-low field (ULF) MRI at 0.13 mT. We have achieved a high longitudinal relaxivity for 18 nm SPION solutions, r1 = 615 s−1 mM−1, which is two orders of magnitude larger than typical commercial Gd-based T1 contrast agents operating at high fields (1.5 T and 3 T). The significantly enhanced r1 value at ultra-low fields is attributed to the coupling of proton spins with SPION magnetic fluctuations (Brownian and Néel) associated with a low frequency peak in the imaginary part of AC susceptibility (χ”). SPION-based T1-weighted ULF MRI has the advantages of enhanced signal, shorter imaging times, and iron-oxide-based nontoxic biocompatible agents. This approach shows promise to become a functional imaging technique, similar to PET, where low spatial resolution is compensated for by important functional information.

UR - http://www.scopus.com/inward/record.url?scp=85051246498&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85051246498&partnerID=8YFLogxK

U2 - 10.1038/s41598-018-30264-5

DO - 10.1038/s41598-018-30264-5

M3 - Article

C2 - 30089881

AN - SCOPUS:85051246498

VL - 8

JO - Scientific Reports

JF - Scientific Reports

SN - 2045-2322

IS - 1

M1 - 11863

ER -