Kinematic analysis and optimization of a novel robot for surgical tool manipulation

Xiaoli Zhang, Carl A Nelson

Research output: Contribution to journalArticle

30 Citations (Scopus)

Abstract

The size and limited dexterity of current surgical robotic systems are factors that limit their usefulness. To improve the level of assimilation of surgical robots in minimally invasive surgery (MIS), a compact, lightweight surgical robotic positioning mechanism with four degrees of freedom (DOFs) (three rotational DOFs and one translation DOF) is proposed in this paper. This spatial mechanism based on a bevel-gear wrist is remotely driven with three rotation axes intersecting at a remote rotation center (the MIS entry port). Forward and inverse kinematics are derived, and these are used for optimizing the mechanism structure given workspace requirements. By evaluating different spherical geared configurations with various link angles and pitch angles, an optimal design is achieved, which performs surgical tool positioning throughout the desired kinematic workspace while occupying a small space bounded by a hemisphere of radius 13.7 cm. This optimized workspace conservatively accounts for collision avoidance between the patient and robot or internally between the robot links. This resultant mechanism is highly compact and yet has the dexterity to cover the extended workspace typically required in telesurgery. It can also be used for tool tracking and skills assessment. Due to the linear nature of the gearing relationships, it may also be well suited for implementing force feedback for telesurgery.

Original languageEnglish (US)
Article number021003
JournalJournal of Medical Devices, Transactions of the ASME
Volume2
Issue number2
DOIs
StatePublished - Jun 1 2008

Fingerprint

Kinematic Analysis
Workspace
Biomechanical Phenomena
Surgery
Manipulation
Kinematics
Robotics
Minimally Invasive Surgical Procedures
Robot
Robots
Minimally Invasive Surgery
Bevel gears
Inverse kinematics
Optimization
Degree of freedom
Degrees of freedom (mechanics)
Collision avoidance
Positioning
Gears
Wrist

Keywords

  • Forward kinematics
  • Inverse kinematics
  • Minimally invasive surgery
  • Optimization
  • Robotic level-gear wrist
  • Spherical mechanism
  • Surgical robot

ASJC Scopus subject areas

  • Medicine (miscellaneous)
  • Biomedical Engineering

Cite this

Kinematic analysis and optimization of a novel robot for surgical tool manipulation. / Zhang, Xiaoli; Nelson, Carl A.

In: Journal of Medical Devices, Transactions of the ASME, Vol. 2, No. 2, 021003, 01.06.2008.

Research output: Contribution to journalArticle

@article{595352a0f8834b449e3f8b8165f5822b,
title = "Kinematic analysis and optimization of a novel robot for surgical tool manipulation",
abstract = "The size and limited dexterity of current surgical robotic systems are factors that limit their usefulness. To improve the level of assimilation of surgical robots in minimally invasive surgery (MIS), a compact, lightweight surgical robotic positioning mechanism with four degrees of freedom (DOFs) (three rotational DOFs and one translation DOF) is proposed in this paper. This spatial mechanism based on a bevel-gear wrist is remotely driven with three rotation axes intersecting at a remote rotation center (the MIS entry port). Forward and inverse kinematics are derived, and these are used for optimizing the mechanism structure given workspace requirements. By evaluating different spherical geared configurations with various link angles and pitch angles, an optimal design is achieved, which performs surgical tool positioning throughout the desired kinematic workspace while occupying a small space bounded by a hemisphere of radius 13.7 cm. This optimized workspace conservatively accounts for collision avoidance between the patient and robot or internally between the robot links. This resultant mechanism is highly compact and yet has the dexterity to cover the extended workspace typically required in telesurgery. It can also be used for tool tracking and skills assessment. Due to the linear nature of the gearing relationships, it may also be well suited for implementing force feedback for telesurgery.",
keywords = "Forward kinematics, Inverse kinematics, Minimally invasive surgery, Optimization, Robotic level-gear wrist, Spherical mechanism, Surgical robot",
author = "Xiaoli Zhang and Nelson, {Carl A}",
year = "2008",
month = "6",
day = "1",
doi = "10.1115/1.2918740",
language = "English (US)",
volume = "2",
journal = "Journal of Medical Devices, Transactions of the ASME",
issn = "1932-6181",
publisher = "American Society of Mechanical Engineers(ASME)",
number = "2",

}

TY - JOUR

T1 - Kinematic analysis and optimization of a novel robot for surgical tool manipulation

AU - Zhang, Xiaoli

AU - Nelson, Carl A

PY - 2008/6/1

Y1 - 2008/6/1

N2 - The size and limited dexterity of current surgical robotic systems are factors that limit their usefulness. To improve the level of assimilation of surgical robots in minimally invasive surgery (MIS), a compact, lightweight surgical robotic positioning mechanism with four degrees of freedom (DOFs) (three rotational DOFs and one translation DOF) is proposed in this paper. This spatial mechanism based on a bevel-gear wrist is remotely driven with three rotation axes intersecting at a remote rotation center (the MIS entry port). Forward and inverse kinematics are derived, and these are used for optimizing the mechanism structure given workspace requirements. By evaluating different spherical geared configurations with various link angles and pitch angles, an optimal design is achieved, which performs surgical tool positioning throughout the desired kinematic workspace while occupying a small space bounded by a hemisphere of radius 13.7 cm. This optimized workspace conservatively accounts for collision avoidance between the patient and robot or internally between the robot links. This resultant mechanism is highly compact and yet has the dexterity to cover the extended workspace typically required in telesurgery. It can also be used for tool tracking and skills assessment. Due to the linear nature of the gearing relationships, it may also be well suited for implementing force feedback for telesurgery.

AB - The size and limited dexterity of current surgical robotic systems are factors that limit their usefulness. To improve the level of assimilation of surgical robots in minimally invasive surgery (MIS), a compact, lightweight surgical robotic positioning mechanism with four degrees of freedom (DOFs) (three rotational DOFs and one translation DOF) is proposed in this paper. This spatial mechanism based on a bevel-gear wrist is remotely driven with three rotation axes intersecting at a remote rotation center (the MIS entry port). Forward and inverse kinematics are derived, and these are used for optimizing the mechanism structure given workspace requirements. By evaluating different spherical geared configurations with various link angles and pitch angles, an optimal design is achieved, which performs surgical tool positioning throughout the desired kinematic workspace while occupying a small space bounded by a hemisphere of radius 13.7 cm. This optimized workspace conservatively accounts for collision avoidance between the patient and robot or internally between the robot links. This resultant mechanism is highly compact and yet has the dexterity to cover the extended workspace typically required in telesurgery. It can also be used for tool tracking and skills assessment. Due to the linear nature of the gearing relationships, it may also be well suited for implementing force feedback for telesurgery.

KW - Forward kinematics

KW - Inverse kinematics

KW - Minimally invasive surgery

KW - Optimization

KW - Robotic level-gear wrist

KW - Spherical mechanism

KW - Surgical robot

UR - http://www.scopus.com/inward/record.url?scp=50849108196&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=50849108196&partnerID=8YFLogxK

U2 - 10.1115/1.2918740

DO - 10.1115/1.2918740

M3 - Article

VL - 2

JO - Journal of Medical Devices, Transactions of the ASME

JF - Journal of Medical Devices, Transactions of the ASME

SN - 1932-6181

IS - 2

M1 - 021003

ER -