Influence of calibration method on distortion-product otoacoustic emission measurements: I. Test performance

Sienna R. Burke, Abigail R. Rogers, Stephen T Neely, Judy G. Kopun, Hongyang Tan, Michael P Gorga

Research output: Contribution to journalArticle

20 Citations (Scopus)

Abstract

Objectives: Calibration errors in distortion-product otoacoustic emission (DPOAE) measurements because of standing waves cause unpredictable changes in stimulus and DPOAE response level. The purpose of this study was to assess the extent to which these errors affect DPOAE test performance. Standard calibration procedures use sound pressure level (SPL) to determine specified levels. Forward pressure level (FPL) is an alternate calibration method that is less susceptible to standing waves. However, FPL derivation requires prior cavity measurements, which have associated variability. In an attempt to address this variability, four FPL methods were compared with SPL: a reference calibration derived from 25 measurements before all data collection and a daily calibration measurement, both of which were made at body and room temperature. Design: Data were collected from 52 normal-hearing and 103 hearing-impaired subjects. DPOAEs were measured for f2 frequencies ranging from 2 to 8 kHz in half-octave steps, with L2 ranging from -20 to 70 dB SPL (5-dB steps). At each f2, DPOAEs were measured in five calibration conditions: SPL, daily FPL at body temperature (daily body), daily FPL at room temperature (daily room), reference FPL at body temperature (ref body), and reference FPL at room temperature (ref room). Data were used to construct receiver operating characteristic (ROC) curves for each f2, calibration method, and L2. From these curves, areas under the ROC curve (AROC) were estimated. Results: The results of this study are summarized by the following observations: (1) DPOAE test performance was sensitive to stimulus level, regardless of calibration method, with the best test performance observed for moderate stimulus level conditions. (2) An effect of frequency was observed for all calibration methods, with the best test performance at 6 kHz and the worst performance at 8 kHz. (3) At clinically applicable stimulus levels, little difference in test performance among calibration methods was noted across frequencies, except at 8 kHz. At 8 kHz, FPL-based calibration methods provided superior performance compared with the standard SPL calibration. (4) A difference between FPL calibration methods was observed at 8 kHz, with the best test performance occurring for daily calibrations at body temperature. Conclusions: With the exception of 8 kHz, there was little difference in test performance across calibration methods. At 8 kHz, AROCs and specificities for fixed sensitivities indicate that FPL-based calibration methods provide superior performance compared with the standard SPL calibration for clinically relevant levels. Temperature may have an impact on FPL calculations relative to DPOAE test performance. Although the differences in AROC among calibration procedures were not statistically significant, the present results indicate that standing wave errors may impact DPOAE test performance and can be reduced by using FPL, although the largest effects were restricted to 8 kHz.

Original languageEnglish (US)
Pages (from-to)533-545
Number of pages13
JournalEar and hearing
Volume31
Issue number4
DOIs
StatePublished - May 7 2010

Fingerprint

Calibration
Pressure
Body Temperature
ROC Curve
Hearing
Temperature
Area Under Curve

ASJC Scopus subject areas

  • Otorhinolaryngology
  • Speech and Hearing

Cite this

Influence of calibration method on distortion-product otoacoustic emission measurements : I. Test performance. / Burke, Sienna R.; Rogers, Abigail R.; Neely, Stephen T; Kopun, Judy G.; Tan, Hongyang; Gorga, Michael P.

In: Ear and hearing, Vol. 31, No. 4, 07.05.2010, p. 533-545.

Research output: Contribution to journalArticle

Burke, Sienna R. ; Rogers, Abigail R. ; Neely, Stephen T ; Kopun, Judy G. ; Tan, Hongyang ; Gorga, Michael P. / Influence of calibration method on distortion-product otoacoustic emission measurements : I. Test performance. In: Ear and hearing. 2010 ; Vol. 31, No. 4. pp. 533-545.
@article{a715a02b05a4424f9e19f568167357b7,
title = "Influence of calibration method on distortion-product otoacoustic emission measurements: I. Test performance",
abstract = "Objectives: Calibration errors in distortion-product otoacoustic emission (DPOAE) measurements because of standing waves cause unpredictable changes in stimulus and DPOAE response level. The purpose of this study was to assess the extent to which these errors affect DPOAE test performance. Standard calibration procedures use sound pressure level (SPL) to determine specified levels. Forward pressure level (FPL) is an alternate calibration method that is less susceptible to standing waves. However, FPL derivation requires prior cavity measurements, which have associated variability. In an attempt to address this variability, four FPL methods were compared with SPL: a reference calibration derived from 25 measurements before all data collection and a daily calibration measurement, both of which were made at body and room temperature. Design: Data were collected from 52 normal-hearing and 103 hearing-impaired subjects. DPOAEs were measured for f2 frequencies ranging from 2 to 8 kHz in half-octave steps, with L2 ranging from -20 to 70 dB SPL (5-dB steps). At each f2, DPOAEs were measured in five calibration conditions: SPL, daily FPL at body temperature (daily body), daily FPL at room temperature (daily room), reference FPL at body temperature (ref body), and reference FPL at room temperature (ref room). Data were used to construct receiver operating characteristic (ROC) curves for each f2, calibration method, and L2. From these curves, areas under the ROC curve (AROC) were estimated. Results: The results of this study are summarized by the following observations: (1) DPOAE test performance was sensitive to stimulus level, regardless of calibration method, with the best test performance observed for moderate stimulus level conditions. (2) An effect of frequency was observed for all calibration methods, with the best test performance at 6 kHz and the worst performance at 8 kHz. (3) At clinically applicable stimulus levels, little difference in test performance among calibration methods was noted across frequencies, except at 8 kHz. At 8 kHz, FPL-based calibration methods provided superior performance compared with the standard SPL calibration. (4) A difference between FPL calibration methods was observed at 8 kHz, with the best test performance occurring for daily calibrations at body temperature. Conclusions: With the exception of 8 kHz, there was little difference in test performance across calibration methods. At 8 kHz, AROCs and specificities for fixed sensitivities indicate that FPL-based calibration methods provide superior performance compared with the standard SPL calibration for clinically relevant levels. Temperature may have an impact on FPL calculations relative to DPOAE test performance. Although the differences in AROC among calibration procedures were not statistically significant, the present results indicate that standing wave errors may impact DPOAE test performance and can be reduced by using FPL, although the largest effects were restricted to 8 kHz.",
author = "Burke, {Sienna R.} and Rogers, {Abigail R.} and Neely, {Stephen T} and Kopun, {Judy G.} and Hongyang Tan and Gorga, {Michael P}",
year = "2010",
month = "5",
day = "7",
doi = "10.1097/AUD.0b013e3181d86b3d",
language = "English (US)",
volume = "31",
pages = "533--545",
journal = "Ear and Hearing",
issn = "0196-0202",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - Influence of calibration method on distortion-product otoacoustic emission measurements

T2 - I. Test performance

AU - Burke, Sienna R.

AU - Rogers, Abigail R.

AU - Neely, Stephen T

AU - Kopun, Judy G.

AU - Tan, Hongyang

AU - Gorga, Michael P

PY - 2010/5/7

Y1 - 2010/5/7

N2 - Objectives: Calibration errors in distortion-product otoacoustic emission (DPOAE) measurements because of standing waves cause unpredictable changes in stimulus and DPOAE response level. The purpose of this study was to assess the extent to which these errors affect DPOAE test performance. Standard calibration procedures use sound pressure level (SPL) to determine specified levels. Forward pressure level (FPL) is an alternate calibration method that is less susceptible to standing waves. However, FPL derivation requires prior cavity measurements, which have associated variability. In an attempt to address this variability, four FPL methods were compared with SPL: a reference calibration derived from 25 measurements before all data collection and a daily calibration measurement, both of which were made at body and room temperature. Design: Data were collected from 52 normal-hearing and 103 hearing-impaired subjects. DPOAEs were measured for f2 frequencies ranging from 2 to 8 kHz in half-octave steps, with L2 ranging from -20 to 70 dB SPL (5-dB steps). At each f2, DPOAEs were measured in five calibration conditions: SPL, daily FPL at body temperature (daily body), daily FPL at room temperature (daily room), reference FPL at body temperature (ref body), and reference FPL at room temperature (ref room). Data were used to construct receiver operating characteristic (ROC) curves for each f2, calibration method, and L2. From these curves, areas under the ROC curve (AROC) were estimated. Results: The results of this study are summarized by the following observations: (1) DPOAE test performance was sensitive to stimulus level, regardless of calibration method, with the best test performance observed for moderate stimulus level conditions. (2) An effect of frequency was observed for all calibration methods, with the best test performance at 6 kHz and the worst performance at 8 kHz. (3) At clinically applicable stimulus levels, little difference in test performance among calibration methods was noted across frequencies, except at 8 kHz. At 8 kHz, FPL-based calibration methods provided superior performance compared with the standard SPL calibration. (4) A difference between FPL calibration methods was observed at 8 kHz, with the best test performance occurring for daily calibrations at body temperature. Conclusions: With the exception of 8 kHz, there was little difference in test performance across calibration methods. At 8 kHz, AROCs and specificities for fixed sensitivities indicate that FPL-based calibration methods provide superior performance compared with the standard SPL calibration for clinically relevant levels. Temperature may have an impact on FPL calculations relative to DPOAE test performance. Although the differences in AROC among calibration procedures were not statistically significant, the present results indicate that standing wave errors may impact DPOAE test performance and can be reduced by using FPL, although the largest effects were restricted to 8 kHz.

AB - Objectives: Calibration errors in distortion-product otoacoustic emission (DPOAE) measurements because of standing waves cause unpredictable changes in stimulus and DPOAE response level. The purpose of this study was to assess the extent to which these errors affect DPOAE test performance. Standard calibration procedures use sound pressure level (SPL) to determine specified levels. Forward pressure level (FPL) is an alternate calibration method that is less susceptible to standing waves. However, FPL derivation requires prior cavity measurements, which have associated variability. In an attempt to address this variability, four FPL methods were compared with SPL: a reference calibration derived from 25 measurements before all data collection and a daily calibration measurement, both of which were made at body and room temperature. Design: Data were collected from 52 normal-hearing and 103 hearing-impaired subjects. DPOAEs were measured for f2 frequencies ranging from 2 to 8 kHz in half-octave steps, with L2 ranging from -20 to 70 dB SPL (5-dB steps). At each f2, DPOAEs were measured in five calibration conditions: SPL, daily FPL at body temperature (daily body), daily FPL at room temperature (daily room), reference FPL at body temperature (ref body), and reference FPL at room temperature (ref room). Data were used to construct receiver operating characteristic (ROC) curves for each f2, calibration method, and L2. From these curves, areas under the ROC curve (AROC) were estimated. Results: The results of this study are summarized by the following observations: (1) DPOAE test performance was sensitive to stimulus level, regardless of calibration method, with the best test performance observed for moderate stimulus level conditions. (2) An effect of frequency was observed for all calibration methods, with the best test performance at 6 kHz and the worst performance at 8 kHz. (3) At clinically applicable stimulus levels, little difference in test performance among calibration methods was noted across frequencies, except at 8 kHz. At 8 kHz, FPL-based calibration methods provided superior performance compared with the standard SPL calibration. (4) A difference between FPL calibration methods was observed at 8 kHz, with the best test performance occurring for daily calibrations at body temperature. Conclusions: With the exception of 8 kHz, there was little difference in test performance across calibration methods. At 8 kHz, AROCs and specificities for fixed sensitivities indicate that FPL-based calibration methods provide superior performance compared with the standard SPL calibration for clinically relevant levels. Temperature may have an impact on FPL calculations relative to DPOAE test performance. Although the differences in AROC among calibration procedures were not statistically significant, the present results indicate that standing wave errors may impact DPOAE test performance and can be reduced by using FPL, although the largest effects were restricted to 8 kHz.

UR - http://www.scopus.com/inward/record.url?scp=77955852611&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77955852611&partnerID=8YFLogxK

U2 - 10.1097/AUD.0b013e3181d86b3d

DO - 10.1097/AUD.0b013e3181d86b3d

M3 - Article

C2 - 20458246

AN - SCOPUS:77955852611

VL - 31

SP - 533

EP - 545

JO - Ear and Hearing

JF - Ear and Hearing

SN - 0196-0202

IS - 4

ER -