In vitro and in situ characterization of arthroscopic loop security and knot security of braided polyblend sutures: a biomechanical study

Lucas C. Armstrong, Alexander Chong, Ryan W. Livermore, Daniel J. Prohaska, Amanda N. Doyon, Paul H. Wooley

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

We conducted a study to evaluate biomechanical performance during destructive testing of several different suture materials in various arthroscopic knot configurations under both in vitro and in situ conditions. Surgeons of different levels of experience tied the knots. Three different arthroscopic knots (static surgeon's, Weston, Tennessee slider) with 3 reverse half-hitches on alternating posts were tested using Fiberwire, ForceFiber, Orthocord, and Ultrabraid suture materials under both in vitro and in situ (blood plasma at 37°C) conditions. Three surgeons of different experience levels tied the knots on a post 30 mm in circumference. A single load-to-failure test was performed. There were no significant in vitro-in situ differences for Ultrabraid in the different knot configurations or with the different experience levels. Surgeon B (intermediate experience) showed no significant differences between test conditions for any knot configuration or suture material. With Tennessee slider knots, surgeon C (least experience) showed significantly lower clinical failure load under both test conditions and had a higher percentage of complete knot slippage. Surgeon B had no knot slippage with use of Fiberwire. Both the aqueous environment and the surgeon's familiarity with certain knots have an effect on knot security.

Original languageEnglish (US)
Pages (from-to)176-182
Number of pages7
JournalAmerican journal of orthopedics (Belle Mead, N.J.)
Volume44
Issue number4
StatePublished - Apr 1 2015

Fingerprint

Sutures
In Vitro Techniques
Surgeons

ASJC Scopus subject areas

  • Medicine(all)

Cite this

In vitro and in situ characterization of arthroscopic loop security and knot security of braided polyblend sutures : a biomechanical study. / Armstrong, Lucas C.; Chong, Alexander; Livermore, Ryan W.; Prohaska, Daniel J.; Doyon, Amanda N.; Wooley, Paul H.

In: American journal of orthopedics (Belle Mead, N.J.), Vol. 44, No. 4, 01.04.2015, p. 176-182.

Research output: Contribution to journalArticle

Armstrong, Lucas C. ; Chong, Alexander ; Livermore, Ryan W. ; Prohaska, Daniel J. ; Doyon, Amanda N. ; Wooley, Paul H. / In vitro and in situ characterization of arthroscopic loop security and knot security of braided polyblend sutures : a biomechanical study. In: American journal of orthopedics (Belle Mead, N.J.). 2015 ; Vol. 44, No. 4. pp. 176-182.
@article{687302fae3b44f2e8ecb488e5c47041d,
title = "In vitro and in situ characterization of arthroscopic loop security and knot security of braided polyblend sutures: a biomechanical study",
abstract = "We conducted a study to evaluate biomechanical performance during destructive testing of several different suture materials in various arthroscopic knot configurations under both in vitro and in situ conditions. Surgeons of different levels of experience tied the knots. Three different arthroscopic knots (static surgeon's, Weston, Tennessee slider) with 3 reverse half-hitches on alternating posts were tested using Fiberwire, ForceFiber, Orthocord, and Ultrabraid suture materials under both in vitro and in situ (blood plasma at 37°C) conditions. Three surgeons of different experience levels tied the knots on a post 30 mm in circumference. A single load-to-failure test was performed. There were no significant in vitro-in situ differences for Ultrabraid in the different knot configurations or with the different experience levels. Surgeon B (intermediate experience) showed no significant differences between test conditions for any knot configuration or suture material. With Tennessee slider knots, surgeon C (least experience) showed significantly lower clinical failure load under both test conditions and had a higher percentage of complete knot slippage. Surgeon B had no knot slippage with use of Fiberwire. Both the aqueous environment and the surgeon's familiarity with certain knots have an effect on knot security.",
author = "Armstrong, {Lucas C.} and Alexander Chong and Livermore, {Ryan W.} and Prohaska, {Daniel J.} and Doyon, {Amanda N.} and Wooley, {Paul H.}",
year = "2015",
month = "4",
day = "1",
language = "English (US)",
volume = "44",
pages = "176--182",
journal = "American Journal of Orthopedics",
issn = "1078-4519",
publisher = "Quadrant Healthcom Inc.",
number = "4",

}

TY - JOUR

T1 - In vitro and in situ characterization of arthroscopic loop security and knot security of braided polyblend sutures

T2 - a biomechanical study

AU - Armstrong, Lucas C.

AU - Chong, Alexander

AU - Livermore, Ryan W.

AU - Prohaska, Daniel J.

AU - Doyon, Amanda N.

AU - Wooley, Paul H.

PY - 2015/4/1

Y1 - 2015/4/1

N2 - We conducted a study to evaluate biomechanical performance during destructive testing of several different suture materials in various arthroscopic knot configurations under both in vitro and in situ conditions. Surgeons of different levels of experience tied the knots. Three different arthroscopic knots (static surgeon's, Weston, Tennessee slider) with 3 reverse half-hitches on alternating posts were tested using Fiberwire, ForceFiber, Orthocord, and Ultrabraid suture materials under both in vitro and in situ (blood plasma at 37°C) conditions. Three surgeons of different experience levels tied the knots on a post 30 mm in circumference. A single load-to-failure test was performed. There were no significant in vitro-in situ differences for Ultrabraid in the different knot configurations or with the different experience levels. Surgeon B (intermediate experience) showed no significant differences between test conditions for any knot configuration or suture material. With Tennessee slider knots, surgeon C (least experience) showed significantly lower clinical failure load under both test conditions and had a higher percentage of complete knot slippage. Surgeon B had no knot slippage with use of Fiberwire. Both the aqueous environment and the surgeon's familiarity with certain knots have an effect on knot security.

AB - We conducted a study to evaluate biomechanical performance during destructive testing of several different suture materials in various arthroscopic knot configurations under both in vitro and in situ conditions. Surgeons of different levels of experience tied the knots. Three different arthroscopic knots (static surgeon's, Weston, Tennessee slider) with 3 reverse half-hitches on alternating posts were tested using Fiberwire, ForceFiber, Orthocord, and Ultrabraid suture materials under both in vitro and in situ (blood plasma at 37°C) conditions. Three surgeons of different experience levels tied the knots on a post 30 mm in circumference. A single load-to-failure test was performed. There were no significant in vitro-in situ differences for Ultrabraid in the different knot configurations or with the different experience levels. Surgeon B (intermediate experience) showed no significant differences between test conditions for any knot configuration or suture material. With Tennessee slider knots, surgeon C (least experience) showed significantly lower clinical failure load under both test conditions and had a higher percentage of complete knot slippage. Surgeon B had no knot slippage with use of Fiberwire. Both the aqueous environment and the surgeon's familiarity with certain knots have an effect on knot security.

UR - http://www.scopus.com/inward/record.url?scp=84953343522&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84953343522&partnerID=8YFLogxK

M3 - Article

C2 - 25844588

AN - SCOPUS:84953343522

VL - 44

SP - 176

EP - 182

JO - American Journal of Orthopedics

JF - American Journal of Orthopedics

SN - 1078-4519

IS - 4

ER -