Percutaneous transtracheal ventilation using a large gauge intravenoustype catheter can be used successfully in the setting of complete upper airway obstruction in animals. In this study, using a large animal model, satisfactory oxygenation and ventilation was achieved by inversely varying the catheter size and the inspiration to expiration ratio (I:E). Specifically, 30 to 63 kg ruminants with an obstructed upper airway were resuscitated for 30 minutes from a hypoxic, hypercarbic, and acidotic state using 12- and 14-gauge catheters connected to a 50 psl oxygen source via a two-way valve with an I:E of 1:4 and 1:9 seconds, respectively. Shorter expiratory time or increased inspiratory time with these intravenous catheters resulted in significant hemodynamic compromise, barotrauma, inadequate carbon dioxide elimination, acidemia, and frequent death.

Original languageEnglish (US)
Pages (from-to)184-189
Number of pages6
JournalAmerican Journal of Emergency Medicine
Volume8
Issue number3
DOIs
StatePublished - Jan 1 1990

Fingerprint

Ventilation
Catheters
Pressure
Barotrauma
Ruminants
Airway Obstruction
Carbon Dioxide
Animal Models
Hemodynamics
Oxygen

Keywords

  • Percutaneous transtracheal ventilation
  • barotrauma
  • inspiration: expiratory ratio
  • upper airway obstruction

ASJC Scopus subject areas

  • Emergency Medicine

Cite this

High pressure percutaneous transtracheal ventilation : The use of large gauge intravenous-type catheters in the totally obstructed airway. / Stothert Jr, Joseph Clarke; Stout, Mark J.; Lewis, Lawrence M.; Keltner, Raymond M.

In: American Journal of Emergency Medicine, Vol. 8, No. 3, 01.01.1990, p. 184-189.

Research output: Contribution to journalArticle

@article{3315ec85e438495ca445f8ab36b74a3c,
title = "High pressure percutaneous transtracheal ventilation: The use of large gauge intravenous-type catheters in the totally obstructed airway",
abstract = "Percutaneous transtracheal ventilation using a large gauge intravenoustype catheter can be used successfully in the setting of complete upper airway obstruction in animals. In this study, using a large animal model, satisfactory oxygenation and ventilation was achieved by inversely varying the catheter size and the inspiration to expiration ratio (I:E). Specifically, 30 to 63 kg ruminants with an obstructed upper airway were resuscitated for 30 minutes from a hypoxic, hypercarbic, and acidotic state using 12- and 14-gauge catheters connected to a 50 psl oxygen source via a two-way valve with an I:E of 1:4 and 1:9 seconds, respectively. Shorter expiratory time or increased inspiratory time with these intravenous catheters resulted in significant hemodynamic compromise, barotrauma, inadequate carbon dioxide elimination, acidemia, and frequent death.",
keywords = "Percutaneous transtracheal ventilation, barotrauma, inspiration: expiratory ratio, upper airway obstruction",
author = "{Stothert Jr}, {Joseph Clarke} and Stout, {Mark J.} and Lewis, {Lawrence M.} and Keltner, {Raymond M.}",
year = "1990",
month = "1",
day = "1",
doi = "10.1016/0735-6757(90)90318-T",
language = "English (US)",
volume = "8",
pages = "184--189",
journal = "The American journal of emergency medicine",
issn = "0735-6757",
publisher = "W.B. Saunders Ltd",
number = "3",

}

TY - JOUR

T1 - High pressure percutaneous transtracheal ventilation

T2 - The use of large gauge intravenous-type catheters in the totally obstructed airway

AU - Stothert Jr, Joseph Clarke

AU - Stout, Mark J.

AU - Lewis, Lawrence M.

AU - Keltner, Raymond M.

PY - 1990/1/1

Y1 - 1990/1/1

N2 - Percutaneous transtracheal ventilation using a large gauge intravenoustype catheter can be used successfully in the setting of complete upper airway obstruction in animals. In this study, using a large animal model, satisfactory oxygenation and ventilation was achieved by inversely varying the catheter size and the inspiration to expiration ratio (I:E). Specifically, 30 to 63 kg ruminants with an obstructed upper airway were resuscitated for 30 minutes from a hypoxic, hypercarbic, and acidotic state using 12- and 14-gauge catheters connected to a 50 psl oxygen source via a two-way valve with an I:E of 1:4 and 1:9 seconds, respectively. Shorter expiratory time or increased inspiratory time with these intravenous catheters resulted in significant hemodynamic compromise, barotrauma, inadequate carbon dioxide elimination, acidemia, and frequent death.

AB - Percutaneous transtracheal ventilation using a large gauge intravenoustype catheter can be used successfully in the setting of complete upper airway obstruction in animals. In this study, using a large animal model, satisfactory oxygenation and ventilation was achieved by inversely varying the catheter size and the inspiration to expiration ratio (I:E). Specifically, 30 to 63 kg ruminants with an obstructed upper airway were resuscitated for 30 minutes from a hypoxic, hypercarbic, and acidotic state using 12- and 14-gauge catheters connected to a 50 psl oxygen source via a two-way valve with an I:E of 1:4 and 1:9 seconds, respectively. Shorter expiratory time or increased inspiratory time with these intravenous catheters resulted in significant hemodynamic compromise, barotrauma, inadequate carbon dioxide elimination, acidemia, and frequent death.

KW - Percutaneous transtracheal ventilation

KW - barotrauma

KW - inspiration: expiratory ratio

KW - upper airway obstruction

UR - http://www.scopus.com/inward/record.url?scp=0025369511&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0025369511&partnerID=8YFLogxK

U2 - 10.1016/0735-6757(90)90318-T

DO - 10.1016/0735-6757(90)90318-T

M3 - Article

VL - 8

SP - 184

EP - 189

JO - The American journal of emergency medicine

JF - The American journal of emergency medicine

SN - 0735-6757

IS - 3

ER -