Guided growth implant failure is a result of cyclic fatigue: Explant analysis with scanning electron microscopy

Anna E.B. Clement, Ray Vanderby, Matthew A. Halanski, Kenneth J. Noonan

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Background: Guided growth is often used to correct limb deformity and yet implant screw failure in modular systems has been reported. There have been no reports of plate failure and we do not know the exact mode of failure when screws do break. Methods: We report the first published case of a fractured plate in a modular plate and screw construct that was used to correct Blount disease in a child through guided growth. The implants were removed and analyzed for method of failure using scanning electron microscopy. Results: Scanning electron microscopy of the explant confirms that the mode of failure was not a result of static tension from growth. Rather, analysis confirms cyclic fatigue that led to crack propagation across the anterior side of the plate until overload caused complete plate failure. Conclusions: This analysis confirms an in vivo cyclic compression-relaxation of the growth plate presumably to weight-bearing, and that when excessive may lead to implant failure as seen here in this case.

Original languageEnglish (US)
Pages (from-to)e37-e42
JournalJournal of Pediatric Orthopaedics
Volume37
Issue number1
DOIs
StatePublished - Jan 1 2017

Fingerprint

Electron Scanning Microscopy
Fatigue
Growth
Growth Plate
Weight-Bearing
Fourier Analysis
Extremities
Blount disease

Keywords

  • Blount disease
  • guided growth
  • implant failure

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health
  • Orthopedics and Sports Medicine

Cite this

Guided growth implant failure is a result of cyclic fatigue : Explant analysis with scanning electron microscopy. / Clement, Anna E.B.; Vanderby, Ray; Halanski, Matthew A.; Noonan, Kenneth J.

In: Journal of Pediatric Orthopaedics, Vol. 37, No. 1, 01.01.2017, p. e37-e42.

Research output: Contribution to journalArticle

@article{c8febfad2553413c9c78a6b3aa454b77,
title = "Guided growth implant failure is a result of cyclic fatigue: Explant analysis with scanning electron microscopy",
abstract = "Background: Guided growth is often used to correct limb deformity and yet implant screw failure in modular systems has been reported. There have been no reports of plate failure and we do not know the exact mode of failure when screws do break. Methods: We report the first published case of a fractured plate in a modular plate and screw construct that was used to correct Blount disease in a child through guided growth. The implants were removed and analyzed for method of failure using scanning electron microscopy. Results: Scanning electron microscopy of the explant confirms that the mode of failure was not a result of static tension from growth. Rather, analysis confirms cyclic fatigue that led to crack propagation across the anterior side of the plate until overload caused complete plate failure. Conclusions: This analysis confirms an in vivo cyclic compression-relaxation of the growth plate presumably to weight-bearing, and that when excessive may lead to implant failure as seen here in this case.",
keywords = "Blount disease, guided growth, implant failure",
author = "Clement, {Anna E.B.} and Ray Vanderby and Halanski, {Matthew A.} and Noonan, {Kenneth J.}",
year = "2017",
month = "1",
day = "1",
doi = "10.1097/BPO.0000000000000661",
language = "English (US)",
volume = "37",
pages = "e37--e42",
journal = "Journal of Pediatric Orthopaedics",
issn = "0271-6798",
publisher = "Lippincott Williams and Wilkins",
number = "1",

}

TY - JOUR

T1 - Guided growth implant failure is a result of cyclic fatigue

T2 - Explant analysis with scanning electron microscopy

AU - Clement, Anna E.B.

AU - Vanderby, Ray

AU - Halanski, Matthew A.

AU - Noonan, Kenneth J.

PY - 2017/1/1

Y1 - 2017/1/1

N2 - Background: Guided growth is often used to correct limb deformity and yet implant screw failure in modular systems has been reported. There have been no reports of plate failure and we do not know the exact mode of failure when screws do break. Methods: We report the first published case of a fractured plate in a modular plate and screw construct that was used to correct Blount disease in a child through guided growth. The implants were removed and analyzed for method of failure using scanning electron microscopy. Results: Scanning electron microscopy of the explant confirms that the mode of failure was not a result of static tension from growth. Rather, analysis confirms cyclic fatigue that led to crack propagation across the anterior side of the plate until overload caused complete plate failure. Conclusions: This analysis confirms an in vivo cyclic compression-relaxation of the growth plate presumably to weight-bearing, and that when excessive may lead to implant failure as seen here in this case.

AB - Background: Guided growth is often used to correct limb deformity and yet implant screw failure in modular systems has been reported. There have been no reports of plate failure and we do not know the exact mode of failure when screws do break. Methods: We report the first published case of a fractured plate in a modular plate and screw construct that was used to correct Blount disease in a child through guided growth. The implants were removed and analyzed for method of failure using scanning electron microscopy. Results: Scanning electron microscopy of the explant confirms that the mode of failure was not a result of static tension from growth. Rather, analysis confirms cyclic fatigue that led to crack propagation across the anterior side of the plate until overload caused complete plate failure. Conclusions: This analysis confirms an in vivo cyclic compression-relaxation of the growth plate presumably to weight-bearing, and that when excessive may lead to implant failure as seen here in this case.

KW - Blount disease

KW - guided growth

KW - implant failure

UR - http://www.scopus.com/inward/record.url?scp=84946074990&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84946074990&partnerID=8YFLogxK

U2 - 10.1097/BPO.0000000000000661

DO - 10.1097/BPO.0000000000000661

M3 - Article

C2 - 26523701

AN - SCOPUS:84946074990

VL - 37

SP - e37-e42

JO - Journal of Pediatric Orthopaedics

JF - Journal of Pediatric Orthopaedics

SN - 0271-6798

IS - 1

ER -