Epicardial Corkscrew Lead Fracture: An Underreported Cause of Pacing Failure?

W. H. Fleming, L. B. Sarafian, J. W. Yarbrough, John Dale Kugler, P. K. Mooring

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Three children with a pacemaker sustained similar lead fractures within a two-year period. Each fracture occurred in the corkscrew portion of a Medtronic Model 6917 epicardial lead. All patients were boys. The leads had been in place for 12, 45, and 43 months prior to fracture. None of the patients had sustained major trauma. Fracture of the corkscrew lead has been considered an uncommon complication. However, three lead fractures in the corkscrew area in a population of 60 children followed at this institution indicate that this may be a more common cause of pacemaker-system malfunction than previous data suggest. A plea is made for reporting all pacing failures to the manufacturer.

Original languageEnglish (US)
Pages (from-to)535-537
Number of pages3
JournalAnnals of Thoracic Surgery
Volume35
Issue number5
DOIs
StatePublished - Jan 1 1983

Fingerprint

Wounds and Injuries
Population
Lead

ASJC Scopus subject areas

  • Surgery
  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine

Cite this

Epicardial Corkscrew Lead Fracture : An Underreported Cause of Pacing Failure? / Fleming, W. H.; Sarafian, L. B.; Yarbrough, J. W.; Kugler, John Dale; Mooring, P. K.

In: Annals of Thoracic Surgery, Vol. 35, No. 5, 01.01.1983, p. 535-537.

Research output: Contribution to journalArticle

Fleming, W. H. ; Sarafian, L. B. ; Yarbrough, J. W. ; Kugler, John Dale ; Mooring, P. K. / Epicardial Corkscrew Lead Fracture : An Underreported Cause of Pacing Failure?. In: Annals of Thoracic Surgery. 1983 ; Vol. 35, No. 5. pp. 535-537.
@article{d4908946cff74245a91bea1037722da0,
title = "Epicardial Corkscrew Lead Fracture: An Underreported Cause of Pacing Failure?",
abstract = "Three children with a pacemaker sustained similar lead fractures within a two-year period. Each fracture occurred in the corkscrew portion of a Medtronic Model 6917 epicardial lead. All patients were boys. The leads had been in place for 12, 45, and 43 months prior to fracture. None of the patients had sustained major trauma. Fracture of the corkscrew lead has been considered an uncommon complication. However, three lead fractures in the corkscrew area in a population of 60 children followed at this institution indicate that this may be a more common cause of pacemaker-system malfunction than previous data suggest. A plea is made for reporting all pacing failures to the manufacturer.",
author = "Fleming, {W. H.} and Sarafian, {L. B.} and Yarbrough, {J. W.} and Kugler, {John Dale} and Mooring, {P. K.}",
year = "1983",
month = "1",
day = "1",
doi = "10.1016/S0003-4975(10)60428-6",
language = "English (US)",
volume = "35",
pages = "535--537",
journal = "Annals of Thoracic Surgery",
issn = "0003-4975",
publisher = "Elsevier USA",
number = "5",

}

TY - JOUR

T1 - Epicardial Corkscrew Lead Fracture

T2 - An Underreported Cause of Pacing Failure?

AU - Fleming, W. H.

AU - Sarafian, L. B.

AU - Yarbrough, J. W.

AU - Kugler, John Dale

AU - Mooring, P. K.

PY - 1983/1/1

Y1 - 1983/1/1

N2 - Three children with a pacemaker sustained similar lead fractures within a two-year period. Each fracture occurred in the corkscrew portion of a Medtronic Model 6917 epicardial lead. All patients were boys. The leads had been in place for 12, 45, and 43 months prior to fracture. None of the patients had sustained major trauma. Fracture of the corkscrew lead has been considered an uncommon complication. However, three lead fractures in the corkscrew area in a population of 60 children followed at this institution indicate that this may be a more common cause of pacemaker-system malfunction than previous data suggest. A plea is made for reporting all pacing failures to the manufacturer.

AB - Three children with a pacemaker sustained similar lead fractures within a two-year period. Each fracture occurred in the corkscrew portion of a Medtronic Model 6917 epicardial lead. All patients were boys. The leads had been in place for 12, 45, and 43 months prior to fracture. None of the patients had sustained major trauma. Fracture of the corkscrew lead has been considered an uncommon complication. However, three lead fractures in the corkscrew area in a population of 60 children followed at this institution indicate that this may be a more common cause of pacemaker-system malfunction than previous data suggest. A plea is made for reporting all pacing failures to the manufacturer.

UR - http://www.scopus.com/inward/record.url?scp=0020641438&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020641438&partnerID=8YFLogxK

U2 - 10.1016/S0003-4975(10)60428-6

DO - 10.1016/S0003-4975(10)60428-6

M3 - Article

C2 - 6847289

AN - SCOPUS:0020641438

VL - 35

SP - 535

EP - 537

JO - Annals of Thoracic Surgery

JF - Annals of Thoracic Surgery

SN - 0003-4975

IS - 5

ER -