Effects of routine care procedures on transcutaneous oxygen in neonates

A quantitative approach

David Alan Danford, S. Miske, J. Headley, R. M. Nelson

Research output: Contribution to journalArticle

64 Citations (Scopus)

Abstract

The changes in transcutaneous oxygen tension (TcPo2) during 123 observations of 10 care procedures were compared with the intrinsic variability of the TcPo2 at the time of the procedure. Standard deviations of the TcPo2 were measured by planimetry from continuous recording in undisturbed infants during a 10-minute control period immediately before each procedure. The decrement of TcPo2 during the procedure and the change in TcPo2 5 minutes after the procedure had ended were measured. A chest x-ray film had the most striking early hypoxic effect, but all the procedures produced early hypoxia in at least one-quarter of the infants. The significant decreases in oxygenation with particular procedures indicate areas in which changes in techniques might be applied to the benefit of the neonate.

Original languageEnglish (US)
Pages (from-to)20-23
Number of pages4
JournalArchives of Disease in Childhood
Volume58
Issue number1
DOIs
StatePublished - Jan 1 1983

Fingerprint

Newborn Infant
Oxygen
Motion Pictures
Thorax
X-Rays

ASJC Scopus subject areas

  • Pediatrics, Perinatology, and Child Health

Cite this

Effects of routine care procedures on transcutaneous oxygen in neonates : A quantitative approach. / Danford, David Alan; Miske, S.; Headley, J.; Nelson, R. M.

In: Archives of Disease in Childhood, Vol. 58, No. 1, 01.01.1983, p. 20-23.

Research output: Contribution to journalArticle

@article{ec8c7bdca7d94914a7ca9927fb74899d,
title = "Effects of routine care procedures on transcutaneous oxygen in neonates: A quantitative approach",
abstract = "The changes in transcutaneous oxygen tension (TcPo2) during 123 observations of 10 care procedures were compared with the intrinsic variability of the TcPo2 at the time of the procedure. Standard deviations of the TcPo2 were measured by planimetry from continuous recording in undisturbed infants during a 10-minute control period immediately before each procedure. The decrement of TcPo2 during the procedure and the change in TcPo2 5 minutes after the procedure had ended were measured. A chest x-ray film had the most striking early hypoxic effect, but all the procedures produced early hypoxia in at least one-quarter of the infants. The significant decreases in oxygenation with particular procedures indicate areas in which changes in techniques might be applied to the benefit of the neonate.",
author = "Danford, {David Alan} and S. Miske and J. Headley and Nelson, {R. M.}",
year = "1983",
month = "1",
day = "1",
doi = "10.1136/adc.58.1.20",
language = "English (US)",
volume = "58",
pages = "20--23",
journal = "Archives of Disease in Childhood",
issn = "0003-9888",
publisher = "BMJ Publishing Group",
number = "1",

}

TY - JOUR

T1 - Effects of routine care procedures on transcutaneous oxygen in neonates

T2 - A quantitative approach

AU - Danford, David Alan

AU - Miske, S.

AU - Headley, J.

AU - Nelson, R. M.

PY - 1983/1/1

Y1 - 1983/1/1

N2 - The changes in transcutaneous oxygen tension (TcPo2) during 123 observations of 10 care procedures were compared with the intrinsic variability of the TcPo2 at the time of the procedure. Standard deviations of the TcPo2 were measured by planimetry from continuous recording in undisturbed infants during a 10-minute control period immediately before each procedure. The decrement of TcPo2 during the procedure and the change in TcPo2 5 minutes after the procedure had ended were measured. A chest x-ray film had the most striking early hypoxic effect, but all the procedures produced early hypoxia in at least one-quarter of the infants. The significant decreases in oxygenation with particular procedures indicate areas in which changes in techniques might be applied to the benefit of the neonate.

AB - The changes in transcutaneous oxygen tension (TcPo2) during 123 observations of 10 care procedures were compared with the intrinsic variability of the TcPo2 at the time of the procedure. Standard deviations of the TcPo2 were measured by planimetry from continuous recording in undisturbed infants during a 10-minute control period immediately before each procedure. The decrement of TcPo2 during the procedure and the change in TcPo2 5 minutes after the procedure had ended were measured. A chest x-ray film had the most striking early hypoxic effect, but all the procedures produced early hypoxia in at least one-quarter of the infants. The significant decreases in oxygenation with particular procedures indicate areas in which changes in techniques might be applied to the benefit of the neonate.

UR - http://www.scopus.com/inward/record.url?scp=0020693037&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020693037&partnerID=8YFLogxK

U2 - 10.1136/adc.58.1.20

DO - 10.1136/adc.58.1.20

M3 - Article

VL - 58

SP - 20

EP - 23

JO - Archives of Disease in Childhood

JF - Archives of Disease in Childhood

SN - 0003-9888

IS - 1

ER -