Background: Although generally safe, transcatheter patent foramen ovale (PFO) closure typically involves catheter exchanges and manipulations within the left atrium (LA) that may result in complications. We describe the development of a simplified technique for PFO closure, further to modification of our earlier approaches. Methods: Early in our experience, LA angiography and/or balloon sizing of the PFO was routine prior to device closure. Subsequently, the simplified technique described herein became our standard approach. We reviewed immediate and medium term results in 27 consecutive patients who underwent closure in a 24-month time frame. Results: All procedures used general anaesthesia, TEE guidance and CardioSEAL septal occluder. Device delivery was successful in all patients. The median fluoroscopy time was significantly shorter with the simplified technique (median difference 6.7. min, p< 0.05) with a 98.3% median confidence interval (4.5, 19.9. min). There were four complications, all occurring in the early group of patients. There have been no recurrent neurological events or device related adverse events on medium term follow-up. Conclusions: Successful transcatheter PFO closure is feasible without LA angiography or catheter-based balloon sizing, utilising echocardiographic guidance for device selection and placement. Simplification of PFO closure technique could potentially decrease complications and shorten fluoroscopy times.

Original languageEnglish (US)
Pages (from-to)384-387
Number of pages4
JournalHeart Lung and Circulation
Volume18
Issue number6
DOIs
StatePublished - Dec 1 2009

Fingerprint

Patent Foramen Ovale
Heart Atria
Equipment and Supplies
Fluoroscopy
Angiography
Catheters
Septal Occluder Device
General Anesthesia
Confidence Intervals

Keywords

  • Adult congenital heart disease
  • Interventional catheterisation
  • Paediatric cardiology
  • Patent foramen ovale

ASJC Scopus subject areas

  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine

Cite this

Early Experience with a Simplified Technique for Transcatheter Closure of the Patent Foramen Ovale. / Kutty, Shelby; Nykanen, David G.; Zahn, Evan M.

In: Heart Lung and Circulation, Vol. 18, No. 6, 01.12.2009, p. 384-387.

Research output: Contribution to journalArticle

@article{72c58ecf4d054fc28a333c982b04b239,
title = "Early Experience with a Simplified Technique for Transcatheter Closure of the Patent Foramen Ovale",
abstract = "Background: Although generally safe, transcatheter patent foramen ovale (PFO) closure typically involves catheter exchanges and manipulations within the left atrium (LA) that may result in complications. We describe the development of a simplified technique for PFO closure, further to modification of our earlier approaches. Methods: Early in our experience, LA angiography and/or balloon sizing of the PFO was routine prior to device closure. Subsequently, the simplified technique described herein became our standard approach. We reviewed immediate and medium term results in 27 consecutive patients who underwent closure in a 24-month time frame. Results: All procedures used general anaesthesia, TEE guidance and CardioSEAL septal occluder. Device delivery was successful in all patients. The median fluoroscopy time was significantly shorter with the simplified technique (median difference 6.7. min, p< 0.05) with a 98.3{\%} median confidence interval (4.5, 19.9. min). There were four complications, all occurring in the early group of patients. There have been no recurrent neurological events or device related adverse events on medium term follow-up. Conclusions: Successful transcatheter PFO closure is feasible without LA angiography or catheter-based balloon sizing, utilising echocardiographic guidance for device selection and placement. Simplification of PFO closure technique could potentially decrease complications and shorten fluoroscopy times.",
keywords = "Adult congenital heart disease, Interventional catheterisation, Paediatric cardiology, Patent foramen ovale",
author = "Shelby Kutty and Nykanen, {David G.} and Zahn, {Evan M.}",
year = "2009",
month = "12",
day = "1",
doi = "10.1016/j.hlc.2009.08.005",
language = "English (US)",
volume = "18",
pages = "384--387",
journal = "Heart Lung and Circulation",
issn = "1443-9506",
publisher = "Wiley-Blackwell",
number = "6",

}

TY - JOUR

T1 - Early Experience with a Simplified Technique for Transcatheter Closure of the Patent Foramen Ovale

AU - Kutty, Shelby

AU - Nykanen, David G.

AU - Zahn, Evan M.

PY - 2009/12/1

Y1 - 2009/12/1

N2 - Background: Although generally safe, transcatheter patent foramen ovale (PFO) closure typically involves catheter exchanges and manipulations within the left atrium (LA) that may result in complications. We describe the development of a simplified technique for PFO closure, further to modification of our earlier approaches. Methods: Early in our experience, LA angiography and/or balloon sizing of the PFO was routine prior to device closure. Subsequently, the simplified technique described herein became our standard approach. We reviewed immediate and medium term results in 27 consecutive patients who underwent closure in a 24-month time frame. Results: All procedures used general anaesthesia, TEE guidance and CardioSEAL septal occluder. Device delivery was successful in all patients. The median fluoroscopy time was significantly shorter with the simplified technique (median difference 6.7. min, p< 0.05) with a 98.3% median confidence interval (4.5, 19.9. min). There were four complications, all occurring in the early group of patients. There have been no recurrent neurological events or device related adverse events on medium term follow-up. Conclusions: Successful transcatheter PFO closure is feasible without LA angiography or catheter-based balloon sizing, utilising echocardiographic guidance for device selection and placement. Simplification of PFO closure technique could potentially decrease complications and shorten fluoroscopy times.

AB - Background: Although generally safe, transcatheter patent foramen ovale (PFO) closure typically involves catheter exchanges and manipulations within the left atrium (LA) that may result in complications. We describe the development of a simplified technique for PFO closure, further to modification of our earlier approaches. Methods: Early in our experience, LA angiography and/or balloon sizing of the PFO was routine prior to device closure. Subsequently, the simplified technique described herein became our standard approach. We reviewed immediate and medium term results in 27 consecutive patients who underwent closure in a 24-month time frame. Results: All procedures used general anaesthesia, TEE guidance and CardioSEAL septal occluder. Device delivery was successful in all patients. The median fluoroscopy time was significantly shorter with the simplified technique (median difference 6.7. min, p< 0.05) with a 98.3% median confidence interval (4.5, 19.9. min). There were four complications, all occurring in the early group of patients. There have been no recurrent neurological events or device related adverse events on medium term follow-up. Conclusions: Successful transcatheter PFO closure is feasible without LA angiography or catheter-based balloon sizing, utilising echocardiographic guidance for device selection and placement. Simplification of PFO closure technique could potentially decrease complications and shorten fluoroscopy times.

KW - Adult congenital heart disease

KW - Interventional catheterisation

KW - Paediatric cardiology

KW - Patent foramen ovale

UR - http://www.scopus.com/inward/record.url?scp=77449104450&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77449104450&partnerID=8YFLogxK

U2 - 10.1016/j.hlc.2009.08.005

DO - 10.1016/j.hlc.2009.08.005

M3 - Article

C2 - 19783211

AN - SCOPUS:77449104450

VL - 18

SP - 384

EP - 387

JO - Heart Lung and Circulation

JF - Heart Lung and Circulation

SN - 1443-9506

IS - 6

ER -