Determination of size of aortic emboli and embolic load during coronary artery bypass grafting

Denise Barbut, Fun Sun F. Yao, Yi Wen Lo, Ronald Silverman, David N. Hager, Rosario R. Trifiletti, Jeffrey P. Gold

Research output: Contribution to journalArticle

81 Citations (Scopus)

Abstract

Background. Embolic signals have been detected within both the aortic lumen and the intracranial vasculature during coronary artery bypass grafting. Total numbers of these emboli have been reported. The present study examined the size of individual emboli and the total volume of embolization. Methods. Using transesophageal echocardiography, we continuously monitored the aortic lumen of 10 patients undergoing isolated coronary artery bypass grafting. We manually analyzed 720,000 individual echo frames over a 4- minute period after the release of aortic clamps to track and to calculate the volume of 657 individual particles. The embolic load for the entire procedure was calculated from mean volume based on analysis of 1,508 particles. We simultaneously monitored the middle cerebral artery using transcranial Doppler ultrasonography and compared numbers of emboli detected by the two techniques. Results. Particle diameter ranged from 0.3 to 2.9 mm (mean, 0.8 mm), and particle volume from 0.01 to 12.5 mm3 (mean, 0.8 mm3). Twenty-eight percent of particles measured 1 mm or more, 44% measured 0.6 to 1.0 mm, and only 27% measured 0.6 mm or less in diameter. Aortic embolic load for the procedure ranged from 0.6 cm3 to 11.2 cm3 (mean, 3.7 cm3). Estimated cerebral embolic load for the procedure ranged from 60 to 510 mm3 (mean, 276 mm3). The fraction of aortic emboli entering the cerebral circulation was very variable (3.9% to 18.1%). Seventy-six percent of the embolic volume after the release of clamps occurred over a 20-second period. Only 1 patient was encephalopathic perioperatively. This patient had the largest estimated cerebral embolic load (510 mm3) and the second largest aortic embolic load (8.4 cm3). Conclusions. We determined the size of individual intraaortic embolic particles and the total volume of embolization during coronary artery bypass grafting, and found the proportion entering the cerebral circulation to be very variable. The constitution of these particles and the neurologic impairment resulting from such embolization remains to be determined.

Original languageEnglish (US)
Pages (from-to)1262-1265
Number of pages4
JournalAnnals of Thoracic Surgery
Volume63
Issue number5
DOIs
StatePublished - May 1997

Fingerprint

Embolism
Coronary Artery Bypass
Cerebrovascular Circulation
Doppler Transcranial Ultrasonography
Constitution and Bylaws
Transesophageal Echocardiography
Middle Cerebral Artery
Nervous System

ASJC Scopus subject areas

  • Surgery
  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine

Cite this

Determination of size of aortic emboli and embolic load during coronary artery bypass grafting. / Barbut, Denise; Yao, Fun Sun F.; Lo, Yi Wen; Silverman, Ronald; Hager, David N.; Trifiletti, Rosario R.; Gold, Jeffrey P.

In: Annals of Thoracic Surgery, Vol. 63, No. 5, 05.1997, p. 1262-1265.

Research output: Contribution to journalArticle

Barbut, Denise ; Yao, Fun Sun F. ; Lo, Yi Wen ; Silverman, Ronald ; Hager, David N. ; Trifiletti, Rosario R. ; Gold, Jeffrey P. / Determination of size of aortic emboli and embolic load during coronary artery bypass grafting. In: Annals of Thoracic Surgery. 1997 ; Vol. 63, No. 5. pp. 1262-1265.
@article{d16b388732954dcf83afa504b5e5cc75,
title = "Determination of size of aortic emboli and embolic load during coronary artery bypass grafting",
abstract = "Background. Embolic signals have been detected within both the aortic lumen and the intracranial vasculature during coronary artery bypass grafting. Total numbers of these emboli have been reported. The present study examined the size of individual emboli and the total volume of embolization. Methods. Using transesophageal echocardiography, we continuously monitored the aortic lumen of 10 patients undergoing isolated coronary artery bypass grafting. We manually analyzed 720,000 individual echo frames over a 4- minute period after the release of aortic clamps to track and to calculate the volume of 657 individual particles. The embolic load for the entire procedure was calculated from mean volume based on analysis of 1,508 particles. We simultaneously monitored the middle cerebral artery using transcranial Doppler ultrasonography and compared numbers of emboli detected by the two techniques. Results. Particle diameter ranged from 0.3 to 2.9 mm (mean, 0.8 mm), and particle volume from 0.01 to 12.5 mm3 (mean, 0.8 mm3). Twenty-eight percent of particles measured 1 mm or more, 44{\%} measured 0.6 to 1.0 mm, and only 27{\%} measured 0.6 mm or less in diameter. Aortic embolic load for the procedure ranged from 0.6 cm3 to 11.2 cm3 (mean, 3.7 cm3). Estimated cerebral embolic load for the procedure ranged from 60 to 510 mm3 (mean, 276 mm3). The fraction of aortic emboli entering the cerebral circulation was very variable (3.9{\%} to 18.1{\%}). Seventy-six percent of the embolic volume after the release of clamps occurred over a 20-second period. Only 1 patient was encephalopathic perioperatively. This patient had the largest estimated cerebral embolic load (510 mm3) and the second largest aortic embolic load (8.4 cm3). Conclusions. We determined the size of individual intraaortic embolic particles and the total volume of embolization during coronary artery bypass grafting, and found the proportion entering the cerebral circulation to be very variable. The constitution of these particles and the neurologic impairment resulting from such embolization remains to be determined.",
author = "Denise Barbut and Yao, {Fun Sun F.} and Lo, {Yi Wen} and Ronald Silverman and Hager, {David N.} and Trifiletti, {Rosario R.} and Gold, {Jeffrey P.}",
year = "1997",
month = "5",
doi = "10.1016/S0003-4975(97)00196-3",
language = "English (US)",
volume = "63",
pages = "1262--1265",
journal = "Annals of Thoracic Surgery",
issn = "0003-4975",
publisher = "Elsevier USA",
number = "5",

}

TY - JOUR

T1 - Determination of size of aortic emboli and embolic load during coronary artery bypass grafting

AU - Barbut, Denise

AU - Yao, Fun Sun F.

AU - Lo, Yi Wen

AU - Silverman, Ronald

AU - Hager, David N.

AU - Trifiletti, Rosario R.

AU - Gold, Jeffrey P.

PY - 1997/5

Y1 - 1997/5

N2 - Background. Embolic signals have been detected within both the aortic lumen and the intracranial vasculature during coronary artery bypass grafting. Total numbers of these emboli have been reported. The present study examined the size of individual emboli and the total volume of embolization. Methods. Using transesophageal echocardiography, we continuously monitored the aortic lumen of 10 patients undergoing isolated coronary artery bypass grafting. We manually analyzed 720,000 individual echo frames over a 4- minute period after the release of aortic clamps to track and to calculate the volume of 657 individual particles. The embolic load for the entire procedure was calculated from mean volume based on analysis of 1,508 particles. We simultaneously monitored the middle cerebral artery using transcranial Doppler ultrasonography and compared numbers of emboli detected by the two techniques. Results. Particle diameter ranged from 0.3 to 2.9 mm (mean, 0.8 mm), and particle volume from 0.01 to 12.5 mm3 (mean, 0.8 mm3). Twenty-eight percent of particles measured 1 mm or more, 44% measured 0.6 to 1.0 mm, and only 27% measured 0.6 mm or less in diameter. Aortic embolic load for the procedure ranged from 0.6 cm3 to 11.2 cm3 (mean, 3.7 cm3). Estimated cerebral embolic load for the procedure ranged from 60 to 510 mm3 (mean, 276 mm3). The fraction of aortic emboli entering the cerebral circulation was very variable (3.9% to 18.1%). Seventy-six percent of the embolic volume after the release of clamps occurred over a 20-second period. Only 1 patient was encephalopathic perioperatively. This patient had the largest estimated cerebral embolic load (510 mm3) and the second largest aortic embolic load (8.4 cm3). Conclusions. We determined the size of individual intraaortic embolic particles and the total volume of embolization during coronary artery bypass grafting, and found the proportion entering the cerebral circulation to be very variable. The constitution of these particles and the neurologic impairment resulting from such embolization remains to be determined.

AB - Background. Embolic signals have been detected within both the aortic lumen and the intracranial vasculature during coronary artery bypass grafting. Total numbers of these emboli have been reported. The present study examined the size of individual emboli and the total volume of embolization. Methods. Using transesophageal echocardiography, we continuously monitored the aortic lumen of 10 patients undergoing isolated coronary artery bypass grafting. We manually analyzed 720,000 individual echo frames over a 4- minute period after the release of aortic clamps to track and to calculate the volume of 657 individual particles. The embolic load for the entire procedure was calculated from mean volume based on analysis of 1,508 particles. We simultaneously monitored the middle cerebral artery using transcranial Doppler ultrasonography and compared numbers of emboli detected by the two techniques. Results. Particle diameter ranged from 0.3 to 2.9 mm (mean, 0.8 mm), and particle volume from 0.01 to 12.5 mm3 (mean, 0.8 mm3). Twenty-eight percent of particles measured 1 mm or more, 44% measured 0.6 to 1.0 mm, and only 27% measured 0.6 mm or less in diameter. Aortic embolic load for the procedure ranged from 0.6 cm3 to 11.2 cm3 (mean, 3.7 cm3). Estimated cerebral embolic load for the procedure ranged from 60 to 510 mm3 (mean, 276 mm3). The fraction of aortic emboli entering the cerebral circulation was very variable (3.9% to 18.1%). Seventy-six percent of the embolic volume after the release of clamps occurred over a 20-second period. Only 1 patient was encephalopathic perioperatively. This patient had the largest estimated cerebral embolic load (510 mm3) and the second largest aortic embolic load (8.4 cm3). Conclusions. We determined the size of individual intraaortic embolic particles and the total volume of embolization during coronary artery bypass grafting, and found the proportion entering the cerebral circulation to be very variable. The constitution of these particles and the neurologic impairment resulting from such embolization remains to be determined.

UR - http://www.scopus.com/inward/record.url?scp=0031149835&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031149835&partnerID=8YFLogxK

U2 - 10.1016/S0003-4975(97)00196-3

DO - 10.1016/S0003-4975(97)00196-3

M3 - Article

C2 - 9146312

AN - SCOPUS:0031149835

VL - 63

SP - 1262

EP - 1265

JO - Annals of Thoracic Surgery

JF - Annals of Thoracic Surgery

SN - 0003-4975

IS - 5

ER -