Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading

Lucy A. Knight, Saikat Pal, John C. Coleman, Fred Bronson, Hani Haider, Danny L. Levine, Mark Taylor, Paul J. Rullkoetter

Research output: Contribution to journalArticle

107 Citations (Scopus)

Abstract

Pre-clinical experimental wear testing of total knee replacement (TKR) components is an invaluable tool for evaluating new implant designs and materials. However, wear testing can be a lengthy and expensive process, and hence parametric studies evaluating the effects of geometric, loading, or alignment perturbations may at times be cost-prohibitive. The objectives of this study were to develop an adaptive FE method capable of simulating wear of a polyethylene tibial insert and to compare predicted kinematics, weight loss due to wear, and wear depth contours to results from a force-controlled experimental knee simulator. Finite element-based computational wear predictions were performed to 5 million gait cycles using both force- and displacement-controlled inputs. The displacement-controlled inputs, by accurately matching the experimental tibiofemoral motion, provided an evaluation of the simple wear theory. The force-controlled inputs provided an evaluation of the overall numerical method by simultaneously predicting both kinematics and wear. Analysis of the predicted wear convergence behavior indicated that 10 iterations, each representing 500,000 gait cycles, were required to achieve numerical accuracy. Using a wear factor estimated from the literature, the predicted kinematics, polyethylene wear contours, and weight loss were in reasonable agreement with the experimental data, particularly for the stance phase of gait. Although further development of the simplified wear theory is important, the initial predictions are encouraging for future use in design phase implant evaluation. In contrast to the experimental testing which occurred over approximately 2 months, computational wear predictions required only 2 h.

Original languageEnglish (US)
Pages (from-to)1550-1558
Number of pages9
JournalJournal of Biomechanics
Volume40
Issue number7
DOIs
StatePublished - Apr 23 2007

Fingerprint

Knee prostheses
Knee Replacement Arthroplasties
Gait
Biomechanical Phenomena
Wear of materials
Polyethylene
Weight Loss
Knee
Costs and Cost Analysis
Kinematics
Polyethylenes
Testing

Keywords

  • Kinematics
  • Knee mechanics
  • Total knee replacement
  • Wear simulation

ASJC Scopus subject areas

  • Biophysics
  • Orthopedics and Sports Medicine
  • Biomedical Engineering
  • Rehabilitation

Cite this

Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading. / Knight, Lucy A.; Pal, Saikat; Coleman, John C.; Bronson, Fred; Haider, Hani; Levine, Danny L.; Taylor, Mark; Rullkoetter, Paul J.

In: Journal of Biomechanics, Vol. 40, No. 7, 23.04.2007, p. 1550-1558.

Research output: Contribution to journalArticle

Knight, LA, Pal, S, Coleman, JC, Bronson, F, Haider, H, Levine, DL, Taylor, M & Rullkoetter, PJ 2007, 'Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading', Journal of Biomechanics, vol. 40, no. 7, pp. 1550-1558. https://doi.org/10.1016/j.jbiomech.2006.07.027
Knight, Lucy A. ; Pal, Saikat ; Coleman, John C. ; Bronson, Fred ; Haider, Hani ; Levine, Danny L. ; Taylor, Mark ; Rullkoetter, Paul J. / Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading. In: Journal of Biomechanics. 2007 ; Vol. 40, No. 7. pp. 1550-1558.
@article{40481646e463407f80f5c10eab71729e,
title = "Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading",
abstract = "Pre-clinical experimental wear testing of total knee replacement (TKR) components is an invaluable tool for evaluating new implant designs and materials. However, wear testing can be a lengthy and expensive process, and hence parametric studies evaluating the effects of geometric, loading, or alignment perturbations may at times be cost-prohibitive. The objectives of this study were to develop an adaptive FE method capable of simulating wear of a polyethylene tibial insert and to compare predicted kinematics, weight loss due to wear, and wear depth contours to results from a force-controlled experimental knee simulator. Finite element-based computational wear predictions were performed to 5 million gait cycles using both force- and displacement-controlled inputs. The displacement-controlled inputs, by accurately matching the experimental tibiofemoral motion, provided an evaluation of the simple wear theory. The force-controlled inputs provided an evaluation of the overall numerical method by simultaneously predicting both kinematics and wear. Analysis of the predicted wear convergence behavior indicated that 10 iterations, each representing 500,000 gait cycles, were required to achieve numerical accuracy. Using a wear factor estimated from the literature, the predicted kinematics, polyethylene wear contours, and weight loss were in reasonable agreement with the experimental data, particularly for the stance phase of gait. Although further development of the simplified wear theory is important, the initial predictions are encouraging for future use in design phase implant evaluation. In contrast to the experimental testing which occurred over approximately 2 months, computational wear predictions required only 2 h.",
keywords = "Kinematics, Knee mechanics, Total knee replacement, Wear simulation",
author = "Knight, {Lucy A.} and Saikat Pal and Coleman, {John C.} and Fred Bronson and Hani Haider and Levine, {Danny L.} and Mark Taylor and Rullkoetter, {Paul J.}",
year = "2007",
month = "4",
day = "23",
doi = "10.1016/j.jbiomech.2006.07.027",
language = "English (US)",
volume = "40",
pages = "1550--1558",
journal = "Journal of Biomechanics",
issn = "0021-9290",
publisher = "Elsevier Limited",
number = "7",

}

TY - JOUR

T1 - Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading

AU - Knight, Lucy A.

AU - Pal, Saikat

AU - Coleman, John C.

AU - Bronson, Fred

AU - Haider, Hani

AU - Levine, Danny L.

AU - Taylor, Mark

AU - Rullkoetter, Paul J.

PY - 2007/4/23

Y1 - 2007/4/23

N2 - Pre-clinical experimental wear testing of total knee replacement (TKR) components is an invaluable tool for evaluating new implant designs and materials. However, wear testing can be a lengthy and expensive process, and hence parametric studies evaluating the effects of geometric, loading, or alignment perturbations may at times be cost-prohibitive. The objectives of this study were to develop an adaptive FE method capable of simulating wear of a polyethylene tibial insert and to compare predicted kinematics, weight loss due to wear, and wear depth contours to results from a force-controlled experimental knee simulator. Finite element-based computational wear predictions were performed to 5 million gait cycles using both force- and displacement-controlled inputs. The displacement-controlled inputs, by accurately matching the experimental tibiofemoral motion, provided an evaluation of the simple wear theory. The force-controlled inputs provided an evaluation of the overall numerical method by simultaneously predicting both kinematics and wear. Analysis of the predicted wear convergence behavior indicated that 10 iterations, each representing 500,000 gait cycles, were required to achieve numerical accuracy. Using a wear factor estimated from the literature, the predicted kinematics, polyethylene wear contours, and weight loss were in reasonable agreement with the experimental data, particularly for the stance phase of gait. Although further development of the simplified wear theory is important, the initial predictions are encouraging for future use in design phase implant evaluation. In contrast to the experimental testing which occurred over approximately 2 months, computational wear predictions required only 2 h.

AB - Pre-clinical experimental wear testing of total knee replacement (TKR) components is an invaluable tool for evaluating new implant designs and materials. However, wear testing can be a lengthy and expensive process, and hence parametric studies evaluating the effects of geometric, loading, or alignment perturbations may at times be cost-prohibitive. The objectives of this study were to develop an adaptive FE method capable of simulating wear of a polyethylene tibial insert and to compare predicted kinematics, weight loss due to wear, and wear depth contours to results from a force-controlled experimental knee simulator. Finite element-based computational wear predictions were performed to 5 million gait cycles using both force- and displacement-controlled inputs. The displacement-controlled inputs, by accurately matching the experimental tibiofemoral motion, provided an evaluation of the simple wear theory. The force-controlled inputs provided an evaluation of the overall numerical method by simultaneously predicting both kinematics and wear. Analysis of the predicted wear convergence behavior indicated that 10 iterations, each representing 500,000 gait cycles, were required to achieve numerical accuracy. Using a wear factor estimated from the literature, the predicted kinematics, polyethylene wear contours, and weight loss were in reasonable agreement with the experimental data, particularly for the stance phase of gait. Although further development of the simplified wear theory is important, the initial predictions are encouraging for future use in design phase implant evaluation. In contrast to the experimental testing which occurred over approximately 2 months, computational wear predictions required only 2 h.

KW - Kinematics

KW - Knee mechanics

KW - Total knee replacement

KW - Wear simulation

UR - http://www.scopus.com/inward/record.url?scp=34247092956&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34247092956&partnerID=8YFLogxK

U2 - 10.1016/j.jbiomech.2006.07.027

DO - 10.1016/j.jbiomech.2006.07.027

M3 - Article

C2 - 17084405

AN - SCOPUS:34247092956

VL - 40

SP - 1550

EP - 1558

JO - Journal of Biomechanics

JF - Journal of Biomechanics

SN - 0021-9290

IS - 7

ER -