Clinical evaluation of the Terumo Capiox ® FX05 hollow fiber oxygenator with integrated arterial line filter

Joseph Deptula, Melinda Valleley, Kimberly Glogowski, John Detwiler, James M Hammel, Kim F Duncan

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

Perfusion techniques and equipment in pediatric open heart surgery have continued to focus on decreasing prime volumes and lowering surface areas of the cardiopulmonary bypass circuit. While this has improved drastically over the last 20 years, greater demand is being placed on the perfusionist to reduce the deleterious effects of bypass without compromising safety or efficiency. Specifically, manufacturers of disposable perfusion equipment have focused on providing pediatric perfusionists with oxygenators that provide the smallest prime and surface area possible while attempting to maximize performance. Recently, Terumo Cardiovascular has introduced the Capiox® FX05, a neonatal hollow fiber oxygenator that includes an integrated arterial line filter. The FX05 provides a blood flow range of 0.1-1.5 L/min and a low priming volume of 43 mL. Additionally, it is coated with X Coating™, a biocompatible, hydrophilic polymer surface coating that reduces platelet adhesion and protein denaturation. The purpose of this study was to test the FX05 for gas transfer, blood path resistance, and blood handling characteristics in a standardized clinical setting. Heat exchange coefficients were also calculated during the cooling and warming period. Other data analyzed includes bypass circuit prime volumes and initial patient hematocrit along with the total operative homologous blood donor exposures. In summary, the FX05 offers good gas exchange capabilities and a low pressure drop during normal cardiopulmonary bypass parameters along with the safety of an integrated arterial line filter. Furthermore, the FX05 with integrated filter allows a reduction in overall bypass prime volume and surface area while promoting the reduction of homologous blood transfusions, optimizing hemostasis.

Original languageEnglish (US)
Pages (from-to)220-225
Number of pages6
JournalJournal of Extra-Corporeal Technology
Volume41
Issue number4
StatePublished - Dec 1 2009

Fingerprint

Oxygenators
Vascular Access Devices
Cardiopulmonary Bypass
Disposable Equipment
Perfusion
Gases
Pediatrics
Protein Denaturation
Safety
Hemostasis
Blood Donors
Hematocrit
Blood Transfusion
Thoracic Surgery
Polymers
Blood Platelets
Hot Temperature
Pressure
Equipment and Supplies

Keywords

  • Arterial line filter
  • Cardiopulmonary bypass
  • Gas exchange
  • Heat exchange
  • Oxygenator

ASJC Scopus subject areas

  • Medicine (miscellaneous)
  • Health Professions (miscellaneous)
  • Cardiology and Cardiovascular Medicine

Cite this

Clinical evaluation of the Terumo Capiox ® FX05 hollow fiber oxygenator with integrated arterial line filter. / Deptula, Joseph; Valleley, Melinda; Glogowski, Kimberly; Detwiler, John; Hammel, James M; Duncan, Kim F.

In: Journal of Extra-Corporeal Technology, Vol. 41, No. 4, 01.12.2009, p. 220-225.

Research output: Contribution to journalArticle

Deptula, Joseph ; Valleley, Melinda ; Glogowski, Kimberly ; Detwiler, John ; Hammel, James M ; Duncan, Kim F. / Clinical evaluation of the Terumo Capiox ® FX05 hollow fiber oxygenator with integrated arterial line filter. In: Journal of Extra-Corporeal Technology. 2009 ; Vol. 41, No. 4. pp. 220-225.
@article{f6985355cfe4483ab9283ebacf5def06,
title = "Clinical evaluation of the Terumo Capiox {\circledR} FX05 hollow fiber oxygenator with integrated arterial line filter",
abstract = "Perfusion techniques and equipment in pediatric open heart surgery have continued to focus on decreasing prime volumes and lowering surface areas of the cardiopulmonary bypass circuit. While this has improved drastically over the last 20 years, greater demand is being placed on the perfusionist to reduce the deleterious effects of bypass without compromising safety or efficiency. Specifically, manufacturers of disposable perfusion equipment have focused on providing pediatric perfusionists with oxygenators that provide the smallest prime and surface area possible while attempting to maximize performance. Recently, Terumo Cardiovascular has introduced the Capiox{\circledR} FX05, a neonatal hollow fiber oxygenator that includes an integrated arterial line filter. The FX05 provides a blood flow range of 0.1-1.5 L/min and a low priming volume of 43 mL. Additionally, it is coated with X Coating™, a biocompatible, hydrophilic polymer surface coating that reduces platelet adhesion and protein denaturation. The purpose of this study was to test the FX05 for gas transfer, blood path resistance, and blood handling characteristics in a standardized clinical setting. Heat exchange coefficients were also calculated during the cooling and warming period. Other data analyzed includes bypass circuit prime volumes and initial patient hematocrit along with the total operative homologous blood donor exposures. In summary, the FX05 offers good gas exchange capabilities and a low pressure drop during normal cardiopulmonary bypass parameters along with the safety of an integrated arterial line filter. Furthermore, the FX05 with integrated filter allows a reduction in overall bypass prime volume and surface area while promoting the reduction of homologous blood transfusions, optimizing hemostasis.",
keywords = "Arterial line filter, Cardiopulmonary bypass, Gas exchange, Heat exchange, Oxygenator",
author = "Joseph Deptula and Melinda Valleley and Kimberly Glogowski and John Detwiler and Hammel, {James M} and Duncan, {Kim F}",
year = "2009",
month = "12",
day = "1",
language = "English (US)",
volume = "41",
pages = "220--225",
journal = "Journal of Extra-Corporeal Technology",
issn = "0022-1058",
publisher = "American Society of Extra-Corporeal Technology",
number = "4",

}

TY - JOUR

T1 - Clinical evaluation of the Terumo Capiox ® FX05 hollow fiber oxygenator with integrated arterial line filter

AU - Deptula, Joseph

AU - Valleley, Melinda

AU - Glogowski, Kimberly

AU - Detwiler, John

AU - Hammel, James M

AU - Duncan, Kim F

PY - 2009/12/1

Y1 - 2009/12/1

N2 - Perfusion techniques and equipment in pediatric open heart surgery have continued to focus on decreasing prime volumes and lowering surface areas of the cardiopulmonary bypass circuit. While this has improved drastically over the last 20 years, greater demand is being placed on the perfusionist to reduce the deleterious effects of bypass without compromising safety or efficiency. Specifically, manufacturers of disposable perfusion equipment have focused on providing pediatric perfusionists with oxygenators that provide the smallest prime and surface area possible while attempting to maximize performance. Recently, Terumo Cardiovascular has introduced the Capiox® FX05, a neonatal hollow fiber oxygenator that includes an integrated arterial line filter. The FX05 provides a blood flow range of 0.1-1.5 L/min and a low priming volume of 43 mL. Additionally, it is coated with X Coating™, a biocompatible, hydrophilic polymer surface coating that reduces platelet adhesion and protein denaturation. The purpose of this study was to test the FX05 for gas transfer, blood path resistance, and blood handling characteristics in a standardized clinical setting. Heat exchange coefficients were also calculated during the cooling and warming period. Other data analyzed includes bypass circuit prime volumes and initial patient hematocrit along with the total operative homologous blood donor exposures. In summary, the FX05 offers good gas exchange capabilities and a low pressure drop during normal cardiopulmonary bypass parameters along with the safety of an integrated arterial line filter. Furthermore, the FX05 with integrated filter allows a reduction in overall bypass prime volume and surface area while promoting the reduction of homologous blood transfusions, optimizing hemostasis.

AB - Perfusion techniques and equipment in pediatric open heart surgery have continued to focus on decreasing prime volumes and lowering surface areas of the cardiopulmonary bypass circuit. While this has improved drastically over the last 20 years, greater demand is being placed on the perfusionist to reduce the deleterious effects of bypass without compromising safety or efficiency. Specifically, manufacturers of disposable perfusion equipment have focused on providing pediatric perfusionists with oxygenators that provide the smallest prime and surface area possible while attempting to maximize performance. Recently, Terumo Cardiovascular has introduced the Capiox® FX05, a neonatal hollow fiber oxygenator that includes an integrated arterial line filter. The FX05 provides a blood flow range of 0.1-1.5 L/min and a low priming volume of 43 mL. Additionally, it is coated with X Coating™, a biocompatible, hydrophilic polymer surface coating that reduces platelet adhesion and protein denaturation. The purpose of this study was to test the FX05 for gas transfer, blood path resistance, and blood handling characteristics in a standardized clinical setting. Heat exchange coefficients were also calculated during the cooling and warming period. Other data analyzed includes bypass circuit prime volumes and initial patient hematocrit along with the total operative homologous blood donor exposures. In summary, the FX05 offers good gas exchange capabilities and a low pressure drop during normal cardiopulmonary bypass parameters along with the safety of an integrated arterial line filter. Furthermore, the FX05 with integrated filter allows a reduction in overall bypass prime volume and surface area while promoting the reduction of homologous blood transfusions, optimizing hemostasis.

KW - Arterial line filter

KW - Cardiopulmonary bypass

KW - Gas exchange

KW - Heat exchange

KW - Oxygenator

UR - http://www.scopus.com/inward/record.url?scp=75749153220&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=75749153220&partnerID=8YFLogxK

M3 - Article

C2 - 20092076

AN - SCOPUS:75749153220

VL - 41

SP - 220

EP - 225

JO - Journal of Extra-Corporeal Technology

JF - Journal of Extra-Corporeal Technology

SN - 0022-1058

IS - 4

ER -