Bow-tie wobble artifact

Effect of source assembly motion on cone-beam CT

Dandan Zheng, John C. Ford, Jun Lu, Dimitrios Lazos, Geoffrey D. Hugo, Damodar Pokhrel, Lisha Zhang, Jeffrey F. Williamson

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Purpose: To investigate the cause of a bow-tie wobble artifact (BWA) discovered on Varian OBI CBCT images and to develop practical correction strategies.Method and Materials: The dependence of the BWA on phantom geometry, phantom position, specific system, and reconstruction algorithm was investigated. Simulations were conducted to study the dependence of the BWA on scatter and beam hardening corrections. Geometric calibration was performed to rule out other gantry-angle dependent mechanical non-idealities as BWA causes. Air scans were acquired with ball-bearing markers to study the motions of the x-ray head assembly as functions of gantry angle. Based on measurements, we developed hypothesis regarding the BWA cause. Simulations were performed to validate our hypothesis. Two correction strategies were implemented: a measurement-based method, which acquires gantry-dependent normalization projections (NPs); and a model-based method that involves numerically shifting the single-angle NP to compensate for the previously-measured bow-tie-filter (BTF) motion.Results: The BWA has a diameter of ∼15 cm, is centered at the isocenter, and is reproducible independent of phantom, position, system, reconstruction, and standard corrections, but only when the BTF is used. Measurements identified a 2D sinusoidal gantry-angle-dependent motion of the x-ray head assembly, and it was the BTF motion (3 mm amplitude projected onto the detector) resulting an intensity mismatch between the all-angle CBCT projections and a single-angle NP that caused the BWA. Both correction strategies were demonstrated effective.Conclusions: A geometric mismatch between the BTF modulation patterns on CBCT projections and on the NP causes the BWA. The BTF wobble requires additional degrees of freedom in CBCT geometric calibration to characterize.

Original languageEnglish (US)
Pages (from-to)2508-2514
Number of pages7
JournalMedical physics
Volume38
Issue number5
DOIs
StatePublished - Jan 1 2011

Fingerprint

Cone-Beam Computed Tomography
Artifacts
Calibration
Head
X-Rays
Air

Keywords

  • bow-tie-filter
  • cone-beam computed tomography
  • geometric calibration
  • image artifact
  • image-guidance radiation therapy

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Zheng, D., Ford, J. C., Lu, J., Lazos, D., Hugo, G. D., Pokhrel, D., ... Williamson, J. F. (2011). Bow-tie wobble artifact: Effect of source assembly motion on cone-beam CT. Medical physics, 38(5), 2508-2514. https://doi.org/10.1118/1.3582944

Bow-tie wobble artifact : Effect of source assembly motion on cone-beam CT. / Zheng, Dandan; Ford, John C.; Lu, Jun; Lazos, Dimitrios; Hugo, Geoffrey D.; Pokhrel, Damodar; Zhang, Lisha; Williamson, Jeffrey F.

In: Medical physics, Vol. 38, No. 5, 01.01.2011, p. 2508-2514.

Research output: Contribution to journalArticle

Zheng, D, Ford, JC, Lu, J, Lazos, D, Hugo, GD, Pokhrel, D, Zhang, L & Williamson, JF 2011, 'Bow-tie wobble artifact: Effect of source assembly motion on cone-beam CT', Medical physics, vol. 38, no. 5, pp. 2508-2514. https://doi.org/10.1118/1.3582944
Zheng, Dandan ; Ford, John C. ; Lu, Jun ; Lazos, Dimitrios ; Hugo, Geoffrey D. ; Pokhrel, Damodar ; Zhang, Lisha ; Williamson, Jeffrey F. / Bow-tie wobble artifact : Effect of source assembly motion on cone-beam CT. In: Medical physics. 2011 ; Vol. 38, No. 5. pp. 2508-2514.
@article{6849688db1ed41cdae116ee8a942a301,
title = "Bow-tie wobble artifact: Effect of source assembly motion on cone-beam CT",
abstract = "Purpose: To investigate the cause of a bow-tie wobble artifact (BWA) discovered on Varian OBI CBCT images and to develop practical correction strategies.Method and Materials: The dependence of the BWA on phantom geometry, phantom position, specific system, and reconstruction algorithm was investigated. Simulations were conducted to study the dependence of the BWA on scatter and beam hardening corrections. Geometric calibration was performed to rule out other gantry-angle dependent mechanical non-idealities as BWA causes. Air scans were acquired with ball-bearing markers to study the motions of the x-ray head assembly as functions of gantry angle. Based on measurements, we developed hypothesis regarding the BWA cause. Simulations were performed to validate our hypothesis. Two correction strategies were implemented: a measurement-based method, which acquires gantry-dependent normalization projections (NPs); and a model-based method that involves numerically shifting the single-angle NP to compensate for the previously-measured bow-tie-filter (BTF) motion.Results: The BWA has a diameter of ∼15 cm, is centered at the isocenter, and is reproducible independent of phantom, position, system, reconstruction, and standard corrections, but only when the BTF is used. Measurements identified a 2D sinusoidal gantry-angle-dependent motion of the x-ray head assembly, and it was the BTF motion (3 mm amplitude projected onto the detector) resulting an intensity mismatch between the all-angle CBCT projections and a single-angle NP that caused the BWA. Both correction strategies were demonstrated effective.Conclusions: A geometric mismatch between the BTF modulation patterns on CBCT projections and on the NP causes the BWA. The BTF wobble requires additional degrees of freedom in CBCT geometric calibration to characterize.",
keywords = "bow-tie-filter, cone-beam computed tomography, geometric calibration, image artifact, image-guidance radiation therapy",
author = "Dandan Zheng and Ford, {John C.} and Jun Lu and Dimitrios Lazos and Hugo, {Geoffrey D.} and Damodar Pokhrel and Lisha Zhang and Williamson, {Jeffrey F.}",
year = "2011",
month = "1",
day = "1",
doi = "10.1118/1.3582944",
language = "English (US)",
volume = "38",
pages = "2508--2514",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - Bow-tie wobble artifact

T2 - Effect of source assembly motion on cone-beam CT

AU - Zheng, Dandan

AU - Ford, John C.

AU - Lu, Jun

AU - Lazos, Dimitrios

AU - Hugo, Geoffrey D.

AU - Pokhrel, Damodar

AU - Zhang, Lisha

AU - Williamson, Jeffrey F.

PY - 2011/1/1

Y1 - 2011/1/1

N2 - Purpose: To investigate the cause of a bow-tie wobble artifact (BWA) discovered on Varian OBI CBCT images and to develop practical correction strategies.Method and Materials: The dependence of the BWA on phantom geometry, phantom position, specific system, and reconstruction algorithm was investigated. Simulations were conducted to study the dependence of the BWA on scatter and beam hardening corrections. Geometric calibration was performed to rule out other gantry-angle dependent mechanical non-idealities as BWA causes. Air scans were acquired with ball-bearing markers to study the motions of the x-ray head assembly as functions of gantry angle. Based on measurements, we developed hypothesis regarding the BWA cause. Simulations were performed to validate our hypothesis. Two correction strategies were implemented: a measurement-based method, which acquires gantry-dependent normalization projections (NPs); and a model-based method that involves numerically shifting the single-angle NP to compensate for the previously-measured bow-tie-filter (BTF) motion.Results: The BWA has a diameter of ∼15 cm, is centered at the isocenter, and is reproducible independent of phantom, position, system, reconstruction, and standard corrections, but only when the BTF is used. Measurements identified a 2D sinusoidal gantry-angle-dependent motion of the x-ray head assembly, and it was the BTF motion (3 mm amplitude projected onto the detector) resulting an intensity mismatch between the all-angle CBCT projections and a single-angle NP that caused the BWA. Both correction strategies were demonstrated effective.Conclusions: A geometric mismatch between the BTF modulation patterns on CBCT projections and on the NP causes the BWA. The BTF wobble requires additional degrees of freedom in CBCT geometric calibration to characterize.

AB - Purpose: To investigate the cause of a bow-tie wobble artifact (BWA) discovered on Varian OBI CBCT images and to develop practical correction strategies.Method and Materials: The dependence of the BWA on phantom geometry, phantom position, specific system, and reconstruction algorithm was investigated. Simulations were conducted to study the dependence of the BWA on scatter and beam hardening corrections. Geometric calibration was performed to rule out other gantry-angle dependent mechanical non-idealities as BWA causes. Air scans were acquired with ball-bearing markers to study the motions of the x-ray head assembly as functions of gantry angle. Based on measurements, we developed hypothesis regarding the BWA cause. Simulations were performed to validate our hypothesis. Two correction strategies were implemented: a measurement-based method, which acquires gantry-dependent normalization projections (NPs); and a model-based method that involves numerically shifting the single-angle NP to compensate for the previously-measured bow-tie-filter (BTF) motion.Results: The BWA has a diameter of ∼15 cm, is centered at the isocenter, and is reproducible independent of phantom, position, system, reconstruction, and standard corrections, but only when the BTF is used. Measurements identified a 2D sinusoidal gantry-angle-dependent motion of the x-ray head assembly, and it was the BTF motion (3 mm amplitude projected onto the detector) resulting an intensity mismatch between the all-angle CBCT projections and a single-angle NP that caused the BWA. Both correction strategies were demonstrated effective.Conclusions: A geometric mismatch between the BTF modulation patterns on CBCT projections and on the NP causes the BWA. The BTF wobble requires additional degrees of freedom in CBCT geometric calibration to characterize.

KW - bow-tie-filter

KW - cone-beam computed tomography

KW - geometric calibration

KW - image artifact

KW - image-guidance radiation therapy

UR - http://www.scopus.com/inward/record.url?scp=79959551744&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79959551744&partnerID=8YFLogxK

U2 - 10.1118/1.3582944

DO - 10.1118/1.3582944

M3 - Article

VL - 38

SP - 2508

EP - 2514

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -