AFM imaging artifacts due to bacterial cell height and AFM tip geometry

Stephanie B. Velegol, Shaun Pardi, Xu Li, Darrell Velegol, Bruce E. Logan

Research output: Contribution to journalArticle

86 Citations (Scopus)

Abstract

Atomic force microscopy (AFM) has become an important tool for investigating various biological materials, and it is now being applied more routinely for imaging bacteria. By imaging bacteria in water, AFM can provide in-situ images of viable cells and be used to measure interaction forces between the AFM tip (or a colloid probe) and the cell surface. However, the relatively large height and compliance of the bacterium can also lead to imaging artifacts. AFM images of Escherichia coli K12 were consistently found to contain image shadows that were oriented in parallel lines 27° from the direction of the cantilever tilt, regardless of the scan direction. Similar image shadows were also observed for 1 μm diameter polystyrene latex microspheres. Using a simple geometric model for the interaction of the tip and the bacterium, it is demonstrated here that these lines observed for bacteria are image artifacts produced by the pyramidal shape of the tip, the 10° tilt of the cantilever, and the height of the bacterium relative to the size of the tip. Such image artifacts disappear when we image dehydrated bacteria that are lower in height, or bacteria that become damaged and deflated during imaging in water. The interaction of the edge of the tip with the bacterium is also shown to result in inconsistent shapes of force curves unless the force curve is centered on the crest of the rounded bacterial surface.

Original languageEnglish (US)
Pages (from-to)851-857
Number of pages7
JournalLangmuir
Volume19
Issue number3
DOIs
StatePublished - Feb 4 2003

Fingerprint

bacteria
artifacts
Atomic force microscopy
Bacteria
atomic force microscopy
Imaging techniques
Geometry
geometry
cells
Water
Colloids
interactions
curves
Escherichia
latex
Latexes
Microspheres
Biological materials
Escherichia coli
water

ASJC Scopus subject areas

  • Materials Science(all)
  • Condensed Matter Physics
  • Surfaces and Interfaces
  • Spectroscopy
  • Electrochemistry

Cite this

Velegol, S. B., Pardi, S., Li, X., Velegol, D., & Logan, B. E. (2003). AFM imaging artifacts due to bacterial cell height and AFM tip geometry. Langmuir, 19(3), 851-857. https://doi.org/10.1021/la026440g

AFM imaging artifacts due to bacterial cell height and AFM tip geometry. / Velegol, Stephanie B.; Pardi, Shaun; Li, Xu; Velegol, Darrell; Logan, Bruce E.

In: Langmuir, Vol. 19, No. 3, 04.02.2003, p. 851-857.

Research output: Contribution to journalArticle

Velegol, SB, Pardi, S, Li, X, Velegol, D & Logan, BE 2003, 'AFM imaging artifacts due to bacterial cell height and AFM tip geometry', Langmuir, vol. 19, no. 3, pp. 851-857. https://doi.org/10.1021/la026440g
Velegol, Stephanie B. ; Pardi, Shaun ; Li, Xu ; Velegol, Darrell ; Logan, Bruce E. / AFM imaging artifacts due to bacterial cell height and AFM tip geometry. In: Langmuir. 2003 ; Vol. 19, No. 3. pp. 851-857.
@article{c3d7ac23ea6946aa94fe998005f0eef9,
title = "AFM imaging artifacts due to bacterial cell height and AFM tip geometry",
abstract = "Atomic force microscopy (AFM) has become an important tool for investigating various biological materials, and it is now being applied more routinely for imaging bacteria. By imaging bacteria in water, AFM can provide in-situ images of viable cells and be used to measure interaction forces between the AFM tip (or a colloid probe) and the cell surface. However, the relatively large height and compliance of the bacterium can also lead to imaging artifacts. AFM images of Escherichia coli K12 were consistently found to contain image shadows that were oriented in parallel lines 27° from the direction of the cantilever tilt, regardless of the scan direction. Similar image shadows were also observed for 1 μm diameter polystyrene latex microspheres. Using a simple geometric model for the interaction of the tip and the bacterium, it is demonstrated here that these lines observed for bacteria are image artifacts produced by the pyramidal shape of the tip, the 10° tilt of the cantilever, and the height of the bacterium relative to the size of the tip. Such image artifacts disappear when we image dehydrated bacteria that are lower in height, or bacteria that become damaged and deflated during imaging in water. The interaction of the edge of the tip with the bacterium is also shown to result in inconsistent shapes of force curves unless the force curve is centered on the crest of the rounded bacterial surface.",
author = "Velegol, {Stephanie B.} and Shaun Pardi and Xu Li and Darrell Velegol and Logan, {Bruce E.}",
year = "2003",
month = "2",
day = "4",
doi = "10.1021/la026440g",
language = "English (US)",
volume = "19",
pages = "851--857",
journal = "Langmuir",
issn = "0743-7463",
publisher = "American Chemical Society",
number = "3",

}

TY - JOUR

T1 - AFM imaging artifacts due to bacterial cell height and AFM tip geometry

AU - Velegol, Stephanie B.

AU - Pardi, Shaun

AU - Li, Xu

AU - Velegol, Darrell

AU - Logan, Bruce E.

PY - 2003/2/4

Y1 - 2003/2/4

N2 - Atomic force microscopy (AFM) has become an important tool for investigating various biological materials, and it is now being applied more routinely for imaging bacteria. By imaging bacteria in water, AFM can provide in-situ images of viable cells and be used to measure interaction forces between the AFM tip (or a colloid probe) and the cell surface. However, the relatively large height and compliance of the bacterium can also lead to imaging artifacts. AFM images of Escherichia coli K12 were consistently found to contain image shadows that were oriented in parallel lines 27° from the direction of the cantilever tilt, regardless of the scan direction. Similar image shadows were also observed for 1 μm diameter polystyrene latex microspheres. Using a simple geometric model for the interaction of the tip and the bacterium, it is demonstrated here that these lines observed for bacteria are image artifacts produced by the pyramidal shape of the tip, the 10° tilt of the cantilever, and the height of the bacterium relative to the size of the tip. Such image artifacts disappear when we image dehydrated bacteria that are lower in height, or bacteria that become damaged and deflated during imaging in water. The interaction of the edge of the tip with the bacterium is also shown to result in inconsistent shapes of force curves unless the force curve is centered on the crest of the rounded bacterial surface.

AB - Atomic force microscopy (AFM) has become an important tool for investigating various biological materials, and it is now being applied more routinely for imaging bacteria. By imaging bacteria in water, AFM can provide in-situ images of viable cells and be used to measure interaction forces between the AFM tip (or a colloid probe) and the cell surface. However, the relatively large height and compliance of the bacterium can also lead to imaging artifacts. AFM images of Escherichia coli K12 were consistently found to contain image shadows that were oriented in parallel lines 27° from the direction of the cantilever tilt, regardless of the scan direction. Similar image shadows were also observed for 1 μm diameter polystyrene latex microspheres. Using a simple geometric model for the interaction of the tip and the bacterium, it is demonstrated here that these lines observed for bacteria are image artifacts produced by the pyramidal shape of the tip, the 10° tilt of the cantilever, and the height of the bacterium relative to the size of the tip. Such image artifacts disappear when we image dehydrated bacteria that are lower in height, or bacteria that become damaged and deflated during imaging in water. The interaction of the edge of the tip with the bacterium is also shown to result in inconsistent shapes of force curves unless the force curve is centered on the crest of the rounded bacterial surface.

UR - http://www.scopus.com/inward/record.url?scp=0037417910&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037417910&partnerID=8YFLogxK

U2 - 10.1021/la026440g

DO - 10.1021/la026440g

M3 - Article

AN - SCOPUS:0037417910

VL - 19

SP - 851

EP - 857

JO - Langmuir

JF - Langmuir

SN - 0743-7463

IS - 3

ER -