With the development of transgenic and specialized mouse strains, there is an increased need for inhalation exposure systems designed for smaller exposure groups. An inhalation exposure chamber, designed specifically for the exposure of up to 40 mice, was characterized. The chamber was fabricated from 0.32-cm-thick (1/8-in) aluminum sheets with outside dimensions of 61 cm long by 32 cm high by 34 cm deep, resulting in an internal volume of 65 L. Two stainless-steel open-mesh cages, separated by an absorbent barrier, can be stacked within the central portion of the chamber. Access is provided through a gasketed door with a safety-glass face. Tests were performed to determine the chamber leakage rate, degree of mixing, and spatial variation of two aerosols within the chamber. Results indicated that the fractional leakage rate was 0.0003 min-1, well below a reported criterion for an operating chamber. Chamber operation gave similar mixing performance with, or without, use of an interior fan. For aerosols with a mass median aerodynamic diameter (MMAD) of 2.56 μm and 3.14 μm, the spatial variation of particulate matter concentration resulted in coefficients of variation (CVs) of 4.8% and 11.0%, respectively. These CV values are comparable to those obtained from similar studies involving other inhalation exposure chambers.

Original languageEnglish (US)
Pages (from-to)251-263
Number of pages13
JournalInhalation Toxicology
Volume15
Issue number3
DOIs
StatePublished - Mar 1 2003

Fingerprint

Inhalation Exposure
Aerosols
Safety glass
Aluminum sheet
Particulate Matter
Stainless Steel
Eye Protective Devices
Fans
Aerodynamics
Aluminum
Transgenic Mice

ASJC Scopus subject areas

  • Toxicology
  • Health, Toxicology and Mutagenesis

Cite this

A small whole-body exposure chamber for laboratory use. / O'Shaughnessy, Patrick T.; Achutan, Chandran; O'Neill, Marsha E.; Thorne, Peter S.

In: Inhalation Toxicology, Vol. 15, No. 3, 01.03.2003, p. 251-263.

Research output: Contribution to journalArticle

O'Shaughnessy, PT, Achutan, C, O'Neill, ME & Thorne, PS 2003, 'A small whole-body exposure chamber for laboratory use', Inhalation Toxicology, vol. 15, no. 3, pp. 251-263. https://doi.org/10.1080/08958370304504
O'Shaughnessy, Patrick T. ; Achutan, Chandran ; O'Neill, Marsha E. ; Thorne, Peter S. / A small whole-body exposure chamber for laboratory use. In: Inhalation Toxicology. 2003 ; Vol. 15, No. 3. pp. 251-263.
@article{07cb3ea02df04258a93548067c0b5d39,
title = "A small whole-body exposure chamber for laboratory use",
abstract = "With the development of transgenic and specialized mouse strains, there is an increased need for inhalation exposure systems designed for smaller exposure groups. An inhalation exposure chamber, designed specifically for the exposure of up to 40 mice, was characterized. The chamber was fabricated from 0.32-cm-thick (1/8-in) aluminum sheets with outside dimensions of 61 cm long by 32 cm high by 34 cm deep, resulting in an internal volume of 65 L. Two stainless-steel open-mesh cages, separated by an absorbent barrier, can be stacked within the central portion of the chamber. Access is provided through a gasketed door with a safety-glass face. Tests were performed to determine the chamber leakage rate, degree of mixing, and spatial variation of two aerosols within the chamber. Results indicated that the fractional leakage rate was 0.0003 min-1, well below a reported criterion for an operating chamber. Chamber operation gave similar mixing performance with, or without, use of an interior fan. For aerosols with a mass median aerodynamic diameter (MMAD) of 2.56 μm and 3.14 μm, the spatial variation of particulate matter concentration resulted in coefficients of variation (CVs) of 4.8{\%} and 11.0{\%}, respectively. These CV values are comparable to those obtained from similar studies involving other inhalation exposure chambers.",
author = "O'Shaughnessy, {Patrick T.} and Chandran Achutan and O'Neill, {Marsha E.} and Thorne, {Peter S.}",
year = "2003",
month = "3",
day = "1",
doi = "10.1080/08958370304504",
language = "English (US)",
volume = "15",
pages = "251--263",
journal = "Inhalation Toxicology",
issn = "0895-8378",
publisher = "Informa Healthcare",
number = "3",

}

TY - JOUR

T1 - A small whole-body exposure chamber for laboratory use

AU - O'Shaughnessy, Patrick T.

AU - Achutan, Chandran

AU - O'Neill, Marsha E.

AU - Thorne, Peter S.

PY - 2003/3/1

Y1 - 2003/3/1

N2 - With the development of transgenic and specialized mouse strains, there is an increased need for inhalation exposure systems designed for smaller exposure groups. An inhalation exposure chamber, designed specifically for the exposure of up to 40 mice, was characterized. The chamber was fabricated from 0.32-cm-thick (1/8-in) aluminum sheets with outside dimensions of 61 cm long by 32 cm high by 34 cm deep, resulting in an internal volume of 65 L. Two stainless-steel open-mesh cages, separated by an absorbent barrier, can be stacked within the central portion of the chamber. Access is provided through a gasketed door with a safety-glass face. Tests were performed to determine the chamber leakage rate, degree of mixing, and spatial variation of two aerosols within the chamber. Results indicated that the fractional leakage rate was 0.0003 min-1, well below a reported criterion for an operating chamber. Chamber operation gave similar mixing performance with, or without, use of an interior fan. For aerosols with a mass median aerodynamic diameter (MMAD) of 2.56 μm and 3.14 μm, the spatial variation of particulate matter concentration resulted in coefficients of variation (CVs) of 4.8% and 11.0%, respectively. These CV values are comparable to those obtained from similar studies involving other inhalation exposure chambers.

AB - With the development of transgenic and specialized mouse strains, there is an increased need for inhalation exposure systems designed for smaller exposure groups. An inhalation exposure chamber, designed specifically for the exposure of up to 40 mice, was characterized. The chamber was fabricated from 0.32-cm-thick (1/8-in) aluminum sheets with outside dimensions of 61 cm long by 32 cm high by 34 cm deep, resulting in an internal volume of 65 L. Two stainless-steel open-mesh cages, separated by an absorbent barrier, can be stacked within the central portion of the chamber. Access is provided through a gasketed door with a safety-glass face. Tests were performed to determine the chamber leakage rate, degree of mixing, and spatial variation of two aerosols within the chamber. Results indicated that the fractional leakage rate was 0.0003 min-1, well below a reported criterion for an operating chamber. Chamber operation gave similar mixing performance with, or without, use of an interior fan. For aerosols with a mass median aerodynamic diameter (MMAD) of 2.56 μm and 3.14 μm, the spatial variation of particulate matter concentration resulted in coefficients of variation (CVs) of 4.8% and 11.0%, respectively. These CV values are comparable to those obtained from similar studies involving other inhalation exposure chambers.

UR - http://www.scopus.com/inward/record.url?scp=0037352661&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037352661&partnerID=8YFLogxK

U2 - 10.1080/08958370304504

DO - 10.1080/08958370304504

M3 - Article

C2 - 12579456

AN - SCOPUS:0037352661

VL - 15

SP - 251

EP - 263

JO - Inhalation Toxicology

JF - Inhalation Toxicology

SN - 0895-8378

IS - 3

ER -